TL-HEX TruLok Hexapod System, Double Row Footplates, TL-HEX Sterile, All Sizes: 120 MM, 140 MM, 160 MM, 180 MM, 200 MM. The TL-HEX TrueLock Hexapod System (TL-HEX) consists of rings, foot plates and struts to be used in conjunction with the TruLok external fixation system. The system is a metal bone fixation device for stabilization and correction during the normal healing process. The purpose of the TL-HEX System is to provide a solution for deformity correction and keeping the bone segments aligned during the healing process. The device may only be applied by a physician, who is fully responsible for the surgical and post-operative procedures of the system and is able to manage possible mechanical and software limitations. Indications for use for adults and all pediatric subgroups except newborns include: post-traumatic joint contracture which has resulted in loss of range of motion, fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction, open and closed fracture fixation, pseudoarthrosis of long bones, limb lengthening by epiphyseal or metaphyseal distraction, correction of bony or soft tissue deformities, correction of bony of soft tissue defects, joint arthrodesis, infected fractures or non unions.
Published June 13, 2018 · Recall initiated: January 30, 2018
Summary
FDA records medical device Class II recall Z-2063-2018, published June 13, 2018 and initiated January 30, 2018. Recalling firm: Orthofix Srl; Status: Terminated; Reported quantity: 408 units; Lot: a) 120 MM, Product Code 99-56-22000, Lot No. V1374370 (Lot No. laser marked on products V1369866, V1369865, V1369864), Exp. Date 06/08/2019, UDI 18054242512585; b) 140 MM, Product Code 99-56-22020, Lot No. V1374298 (Lot No. laser marked on products V1369870, V1369868, V1369869, V1369872, V1369871), Exp. Date 06/07/2019, UDI 18054242512592; c) 160 MM, Product Code 99-56-22040, Lot No. V1374255 (Lot No. laser marked on products V1370134, V1370136, V1370128, V1370133, V1370132, V1370131), Exp. Date 06/07/2019, Lot No. V1374374 (Lot No. laser marked on products V1370130), Exp. Date 06/08/2019, Lot No. R1065030 (Lot No. laser marked on products V1370134, V1370136, V1370128, V1370133, V1370132, V1370131), Exp. Date 12/16/2021, UDI 18054242512608; d) 180 MM, Product Code 99-56-22060, Lot No. V1374259 (Lot No. laser marked on products V1370581, V1369468, V1369469, V1369467, V1369466, V1369465, V1369471, V1369470, V1370597), Exp. Date 06/07/2019, UDI 18054242512615; e) 200 MM, Product Code 99-56-22080, Lot No. V1374361 (Lot No. laser marked on products V1369462, V1371128, V1371074, V1369463, V1370469), Exp. Date 06/08/2019, UDI 18054242512622; f) 220 MM, Product Code 99-56-22100, Lot No. V1374373 (Lot No. laser marked on products V1369460, V1371250, V1371075), Exp. Date 06/08/2019, UDI 18054242512639. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm received complaints regarding the incompatibility and dimensional interference between the double row footplates manufactured in a limited period of 2014 and struts manufactured after September 2016. The firm determined that certain batches of the double row footplates are not compatible with the current version of the TL-HEX struts.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Orthofix Srl
- Distributor
- Worldwide Distribution - US Distribution within Texas., and to the countries of : Australia, Belgium, Chile, Denmark, France, Germany, Israel, Italy, Japan, South Africa, Spain, Sweden, Switzerland and United Kingdom.
- Lot
- a) 120 MM, Product Code 99-56-22000, Lot No. V1374370 (Lot No. laser marked on products V1369866, V1369865, V1369864), Exp. Date 06/08/2019, UDI 18054242512585; b) 140 MM, Product Code 99-56-22020, Lot No. V1374298 (Lot No. laser marked on products V1369870, V1369868, V1369869, V1369872, V1369871), Exp. Date 06/07/2019, UDI 18054242512592; c) 160 MM, Product Code 99-56-22040, Lot No. V1374255 (Lot No. laser marked on products V1370134, V1370136, V1370128, V1370133, V1370132, V1370131), Exp. Date 06/07/2019, Lot No. V1374374 (Lot No. laser marked on products V1370130), Exp. Date 06/08/2019, Lot No. R1065030 (Lot No. laser marked on products V1370134, V1370136, V1370128, V1370133, V1370132, V1370131), Exp. Date 12/16/2021, UDI 18054242512608; d) 180 MM, Product Code 99-56-22060, Lot No. V1374259 (Lot No. laser marked on products V1370581, V1369468, V1369469, V1369467, V1369466, V1369465, V1369471, V1369470, V1370597), Exp. Date 06/07/2019, UDI 18054242512615; e) 200 MM, Product Code 99-56-22080, Lot No. V1374361 (Lot No. laser marked on products V1369462, V1371128, V1371074, V1369463, V1370469), Exp. Date 06/08/2019, UDI 18054242512622; f) 220 MM, Product Code 99-56-22100, Lot No. V1374373 (Lot No. laser marked on products V1369460, V1371250, V1371075), Exp. Date 06/08/2019, UDI 18054242512639
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Italy
Related recalls
- FDAREF: 99-709245, Fitbone Trochanteric Nail L245mm, D9mm, SterileJune 25, 2025
- FDAORTHOFIX Catalogue Number: ref 1100701, drill bit diameter 4.8 MM LENGTH 280MM, RX ONLY, Non Sterile, UDI:November 20, 2019
- FDAORTHOFIX Catalogue Number: ref 1-1100701,DRILL BIT D.4,8 MM L.280 MM TIN COATED - QUICK CONNECT, RX ONLY, NonNovember 20, 2019
- FDAORTHOFIX Catalogue Number: ref: 99-92501, PREFIX II DIAPHYSEAL COMPLETE KIT STERILE SHORT SCREWS, RX ONLY,November 20, 2019
- FDARX Imola (RX4900) For Professional Use for the quantitative in vitro determination of various clinicalJune 13, 2018
- FDAST-AIA PACK Prolactin; Part Number: 025255 Assay, Reproductive HormoneJune 13, 2018
Questions about this recall
Was TL-HEX TruLok Hexapod System, Double Row Footplates, TL-HEX Sterile, All Sizes: 120 MM, 140 MM, 160 MM, 180 MM, 200 MM. The TL-HEX TrueLock Hexapod System (TL-HEX) consists of rings, foot plates and struts to be used in conjunction with the TruLok external fixation system. The system is a metal bone fixation device for stabilization and correction during the normal healing process. The purpose of the TL-HEX System is to provide a solution for deformity correction and keeping the bone segments aligned during the healing process. The device may only be applied by a physician, who is fully responsible for the surgical and post-operative procedures of the system and is able to manage possible mechanical and software limitations. Indications for use for adults and all pediatric subgroups except newborns include: post-traumatic joint contracture which has resulted in loss of range of motion, fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction, open and closed fracture fixation, pseudoarthrosis of long bones, limb lengthening by epiphyseal or metaphyseal distraction, correction of bony or soft tissue deformities, correction of bony of soft tissue defects, joint arthrodesis, infected fractures or non unions. recalled?
Yes. FDA published recall Z-2063-2018 on June 13, 2018. Reported hazard: The firm received complaints regarding the incompatibility and dimensional interference between the double row footplates manufactured in a limited period of 2014 and struts manufactured after September 2016. The firm determined that certain batches of the double row footplates are not compatible with the current version of the TL-HEX struts.
What should I do if I have this product?
Voluntary: Firm initiated · Letter