Aisys Anesthesia System. Sold under the following product names: AISYS, AISYS 10.X TO 11.X UPGRADE KIT DUTCH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH FOR USA, AISYS 10.X TO 11.X UPGRADE KIT FRENCH, AISYS 10.X TO 11.X UPGRADE KIT GERMAN, AISYS 10.X TO 11.X UPGRADE KIT JAPANESE.
Published November 21, 2018 · Recall initiated: September 27, 2018
Summary
FDA records medical device Class II recall Z-0411-2019, published November 21, 2018 and initiated September 27, 2018. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 4,207 units total; Lot: Model Number, UDI Number, (Serial Number): 2097674-001-US, Not Applicable, (5470345-30); 2097674-001-US, Not Applicable, (5470345-31); 2097674-001-US, Not Applicable, (5470345-32); 1011-9000-000, Not Applicable, (ANAK00774); 1011-9000-000, Not Applicable, (ANAK00775); 1011-9000-000, Not Applicable, (ANAL01201); 1011-9000-000, Not Applicable, (ANAL01234); 1011-9000-000, Not Applicable, (ANAL01235); 1011-9000-000, Not Applicable, (ANAL01236); 1011-9000-000, Not Applicable, (ANAL01237); 1011-9000-000, Not Applicable, (ANAL01238); 1011-9000-000, Not Applicable, (ANAL01240); 1011-9000-000, Not Applicable, (ANAL01494); 1011-9000-000, Not Applicable, (ANAL01495); 1011-9000-000, Not Applicable, (ANAL01551); 1011-9000-000, Not Applicable, (ANAM01023); 1011-9000-000, Not Applicable, (ANAM01024); 1011-9000-000, Not Applicable, (ANAM01025); 1011-9000-000, Not Applicable, (ANAM01026); 1011-9000-000, Not Applicable, (ANAM01027); 1011-9000-000, Not Applicable, (ANAM01104); 1011-9000-000, Not Applicable, (ANAM01105); 1011-9000-000, Not Applicable, (ANAP00438); 1011-9000-000, Not Applicable, (ANAP00439); 1011-9000-000, Not Applicable, (ANAQ01188); 1011-9000-000, Not Applicable, (ANAR05264); 1011-9000-000, Not Applicable, (ANAR05265); 1011-9000-000, Not Applicable, (ANAR05610); 1011-9000-000, Not Applicable, (ANAR05611); 1011-9000-000, Not Applicable, (ANAS00202); 1011-9000-000, Not Applicable, (ANAS00203); 1011-9000-000, Not Applicable, (ANAS00204); 2097674-001-US, Not Applicable, (5479630-1); ANAK00798, Not Applicable (1011-9000-000); ANAK00799, Not Applicable (1011-9000-000); ANAK00803, Not Applicable (1011-9000-000); ANAK00807, Not Applicable (1011-9000-000); ANAK00814, Not Applicable (1011-9000-000); ANAK00839, Not Applicable (1011-9000-000); ANAK00864, Not Applicable (1011-9000-000); 5481223-1, Not Applicable (2097674-001-US); 5481223-10, Not Applicable (2097674-001-US); 5481223-11, Not Applicable (2097674-001-US); 5481223-12, Not Applicable (2097674-001-US); 5481223-2, Not Applicable (2097674-001-US); 5481223-3, Not Applicable (2097674-001-US); 5481223-4, Not Applicable (2097674-001-US); 5481223-5, Not Applicable (2097674-001-US); 5481223-6, Not Applicable (2097674-001-US); 5481223-7, Not Applicable (2097674-001-US); 5481223-8, Not Applicable (2097674-001-US); 5481223-9, Not Applicable (2097674-001-US); ANAP00712, Not Applicable (1011-9000-000); ANAP00713, Not Applicable (1011-9000-000); ANAP00714, Not Applicable (1011-9000-000); ANAN01125, Not Applicable (1011-9000-000); ANAN01138, Not Applicable (1011-9000-000); 4652664-1, Not Applicable (2097674-001-US); 4652664-10, Not Applicable (2097674-001-US); 4652664-11, Not Applicable (2097674-001-US); 4652664-12, Not Applicable (2097674-001-US); 4652664-13, Not Applicable (2097674-001-US); 4652664-2, Not Applicable (2097674-001-US); 4652664-3, Not Applicable (2097674-001-US); 4652664-4, Not Applicable (2097674-001-US); 4652664-5, Not Applicable (2097674-001-US); 4652664-6, Not Applicable (2097674-001-US); 4652664-7, Not Applicable (2097674-001-US); 4652664-8, Not Applicable (2097674-001-US); 4652664-9, Not Applicable (2097674-001-US); ANAK00156, Not Applicable (1011-9000-000); ANAK00157, Not Applicable (1011-9000-000); ANAK00158, Not Applicable (1011-9000-000); ANAK00159, Not Applicable (1011-9000-000); ANAK00511, Not Applicable (1011-9000-000); ANAK00512, Not Applicable (1011-9000-000); ANAK00513, Not Applicable (1011-9000-000); ANAK00514, Not Applicable (1011-9000-000); ANAN01051, Not Applicable (1011-9000-000); ANAN01052, Not Applicable (1011-9000-000); ANAP00545, Not Applicable (1011-9000-000); ANAP00546, Not Applicable (1011-9000-000); ANAP00708, Not Applicable (1011-9000-000); ANAL01400, Not Applicable (1011-9000-000); ANAL01401, Not Applicable (1011-9000-000. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Upgraded Aisys anesthesia devices and Aisys CS2 anesthesia devices with PSVPro Spontaneous Breathing Modes do not provide the PSVPro ventilation mode feature of transitioning to SIMV PCV backup ventilation mode. PSVPro Spontaneous Breathing Modes do not provide the feature of transitioning to SIMV PCV backup ventilation mode for a patient who stops spontaneous breathing if the PSVPro ventilation mode is in use prior to the clinician starting the Cycling Procedure.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare, LLC
- Distributor
- Worldwide distribution - US nationwide in the states of CA, CO, DC, FL, GA, HI, IA, IL, IN, KY, LA, ,MA, MD, MI, MN, MO, MT, NC, NE, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV. Countries of Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Iceland, India, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea (Republic of), Kuwait, Macedonia, Mexico, Netherlands, New Zealand, Nigeria, Norway, Peru, Portugal, Qatar, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam.
- Lot
- Model Number, UDI Number, (Serial Number): 2097674-001-US, Not Applicable, (5470345-30); 2097674-001-US, Not Applicable, (5470345-31); 2097674-001-US, Not Applicable, (5470345-32); 1011-9000-000, Not Applicable, (ANAK00774); 1011-9000-000, Not Applicable, (ANAK00775); 1011-9000-000, Not Applicable, (ANAL01201); 1011-9000-000, Not Applicable, (ANAL01234); 1011-9000-000, Not Applicable, (ANAL01235); 1011-9000-000, Not Applicable, (ANAL01236); 1011-9000-000, Not Applicable, (ANAL01237); 1011-9000-000, Not Applicable, (ANAL01238); 1011-9000-000, Not Applicable, (ANAL01240); 1011-9000-000, Not Applicable, (ANAL01494); 1011-9000-000, Not Applicable, (ANAL01495); 1011-9000-000, Not Applicable, (ANAL01551); 1011-9000-000, Not Applicable, (ANAM01023); 1011-9000-000, Not Applicable, (ANAM01024); 1011-9000-000, Not Applicable, (ANAM01025); 1011-9000-000, Not Applicable, (ANAM01026); 1011-9000-000, Not Applicable, (ANAM01027); 1011-9000-000, Not Applicable, (ANAM01104); 1011-9000-000, Not Applicable, (ANAM01105); 1011-9000-000, Not Applicable, (ANAP00438); 1011-9000-000, Not Applicable, (ANAP00439); 1011-9000-000, Not Applicable, (ANAQ01188); 1011-9000-000, Not Applicable, (ANAR05264); 1011-9000-000, Not Applicable, (ANAR05265); 1011-9000-000, Not Applicable, (ANAR05610); 1011-9000-000, Not Applicable, (ANAR05611); 1011-9000-000, Not Applicable, (ANAS00202); 1011-9000-000, Not Applicable, (ANAS00203); 1011-9000-000, Not Applicable, (ANAS00204); 2097674-001-US, Not Applicable, (5479630-1); ANAK00798, Not Applicable (1011-9000-000); ANAK00799, Not Applicable (1011-9000-000); ANAK00803, Not Applicable (1011-9000-000); ANAK00807, Not Applicable (1011-9000-000); ANAK00814, Not Applicable (1011-9000-000); ANAK00839, Not Applicable (1011-9000-000); ANAK00864, Not Applicable (1011-9000-000); 5481223-1, Not Applicable (2097674-001-US); 5481223-10, Not Applicable (2097674-001-US); 5481223-11, Not Applicable (2097674-001-US); 5481223-12, Not Applicable (2097674-001-US); 5481223-2, Not Applicable (2097674-001-US); 5481223-3, Not Applicable (2097674-001-US); 5481223-4, Not Applicable (2097674-001-US); 5481223-5, Not Applicable (2097674-001-US); 5481223-6, Not Applicable (2097674-001-US); 5481223-7, Not Applicable (2097674-001-US); 5481223-8, Not Applicable (2097674-001-US); 5481223-9, Not Applicable (2097674-001-US); ANAP00712, Not Applicable (1011-9000-000); ANAP00713, Not Applicable (1011-9000-000); ANAP00714, Not Applicable (1011-9000-000); ANAN01125, Not Applicable (1011-9000-000); ANAN01138, Not Applicable (1011-9000-000); 4652664-1, Not Applicable (2097674-001-US); 4652664-10, Not Applicable (2097674-001-US); 4652664-11, Not Applicable (2097674-001-US); 4652664-12, Not Applicable (2097674-001-US); 4652664-13, Not Applicable (2097674-001-US); 4652664-2, Not Applicable (2097674-001-US); 4652664-3, Not Applicable (2097674-001-US); 4652664-4, Not Applicable (2097674-001-US); 4652664-5, Not Applicable (2097674-001-US); 4652664-6, Not Applicable (2097674-001-US); 4652664-7, Not Applicable (2097674-001-US); 4652664-8, Not Applicable (2097674-001-US); 4652664-9, Not Applicable (2097674-001-US); ANAK00156, Not Applicable (1011-9000-000); ANAK00157, Not Applicable (1011-9000-000); ANAK00158, Not Applicable (1011-9000-000); ANAK00159, Not Applicable (1011-9000-000); ANAK00511, Not Applicable (1011-9000-000); ANAK00512, Not Applicable (1011-9000-000); ANAK00513, Not Applicable (1011-9000-000); ANAK00514, Not Applicable (1011-9000-000); ANAN01051, Not Applicable (1011-9000-000); ANAN01052, Not Applicable (1011-9000-000); ANAP00545, Not Applicable (1011-9000-000); ANAP00546, Not Applicable (1011-9000-000); ANAP00708, Not Applicable (1011-9000-000); ANAL01400, Not Applicable (1011-9000-000); ANAL01401, Not Applicable (1011-9000-000
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACentricity PACS Software Version 7.0 SP0.0.4.7August 3, 2022
- FDASIGNA Premier magnetic resonance scanner, model 5748519.July 27, 2022
- FDAGE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images andJuly 13, 2022
- FDABack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019July 6, 2022
- FDACurry powder packaged under bulk, retail, and food service labels. Bulk: CURRY POWDER Net Weight 55 LBS,November 21, 2018
- FDAAisys CS2 Anesthesia System. Sold under the following product names: AISYS, AISYS 10.X TO 11.X UPGRADE KITNovember 21, 2018
Questions about this recall
Was Aisys Anesthesia System. Sold under the following product names: AISYS, AISYS 10.X TO 11.X UPGRADE KIT DUTCH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH FOR USA, AISYS 10.X TO 11.X UPGRADE KIT FRENCH, AISYS 10.X TO 11.X UPGRADE KIT GERMAN, AISYS 10.X TO 11.X UPGRADE KIT JAPANESE. recalled?
Yes. FDA published recall Z-0411-2019 on November 21, 2018. Reported hazard: Upgraded Aisys anesthesia devices and Aisys CS2 anesthesia devices with PSVPro Spontaneous Breathing Modes do not provide the PSVPro ventilation mode feature of transitioning to SIMV PCV backup ventilation mode. PSVPro Spontaneous Breathing Modes do not provide the feature of transitioning to SIMV PCV backup ventilation mode for a patient who stops spontaneous breathing if the PSVPro ventilation mode is in use prior to the clinician starting the Cycling Procedure.
What should I do if I have this product?
Voluntary: Firm initiated · Letter