Aisys CS2 Anesthesia System. Sold under the following product names: AISYS, AISYS 10.X TO 11.X UPGRADE KIT DUTCH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH FOR USA, AISYS 10.X TO 11.X UPGRADE KIT FRENCH, AISYS 10.X TO 11.X UPGRADE KIT GERMAN, AISYS 10.X TO 11.X UPGRADE KIT JAPANESE. Product Usage: The GE Datex-Ohmeda Aisys Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). The device is intended for volume or pressure control ventilation. The Aisys is not suitable for use in a MRI environment.
Published November 21, 2018 · Recall initiated: September 27, 2018
Summary
FDA records medical device Class II recall Z-0412-2019, published November 21, 2018 and initiated September 27, 2018. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 4,207 units total; Lot: Model Number, UDI Number, (Serial Number): 1011-9050-000, 010084068210229221APWX0132811180608, (APWX01328); 1011-9050-000, 010084068210229221APWX0048411180208, (APWX00484); 1011-9050-000, 010084068210229221APWW0214911171214, (APWW02149); 1011-9050-000, 10084068210229221APWW0215011171214, (APWW02150); 1011-9050-000, 010084068210229221APWX0026111180118, (APWX00261); 1011-9050-000, 010084068210229221APWX0026211180118, (APWX00262); 1011-9050-000, 010084068210229221APWX0140911180618, (APWX01409); 1011-9050-000, 010084068210229221APWX0141011180618, (APWX01410); 1011-9050-000, 010084068210229221APWX0141111180618, (APWX01411); 1011-9050-000, 010084068210229221APWX0141211180618, (APWX01412); 1011-9050-000, 010084068210229221APWX0141311180618, (APWX01413); 1011-9050-000, 010084068210229221APWX0141411180618, (APWX01414); 1011-9050-000, 010084068210229221APWX0141511180618, (APWX01415); 1011-9050-000, 010084068210229221APWX0141611180618, (APWX01416); 1011-9050-000, 010084068210229221APWX0141711180618, (APWX01417); 1011-9050-000, 010084068210229221APWX0141811180618, (APWX01418); 1011-9050-000, 010084068210229221APWX0141911180618, (APWX01419); 1011-9050-000, 010084068210229221APWX0142011180618, (APWX01420); 1011-9050-000, 010084068210229221APWX0142111180618, (APWX01421); 1011-9050-000, 010084068210229221APWX0142211180618, (APWX01422); 1011-9050-000, 010084068210229221APWX0026311180118, (APWX00263); 1011-9050-000, 010084068210229221APWX0026411180118, (APWX00264); 1011-9050-000, 010084068210229221APWX0026511180118, (APWX00265); 1011-9050-000, 010084068210229221APWW0214011171214, (APWW02140); 1011-9050-000, 010084068210229221APWW0215611171214, (APWW02156); 1011-9050-000, 010084068210229221APWW0215711171214, (APWW02157); 1011-9050-000, 010084068210229221APWX0158911180703, (APWX01589); 1011-9050-000, 010084068210229221APWX0159011180703, (APWX01590); 1011-9050-000, 010084068210229221APWX0159111180703, (APWX01591); 1011-9050-000, 010084068210229221APWX0159211180703, (APWX01592); 1011-9050-000, 010084068210229221APWX0159311180703, (APWX01593); 1011-9050-000, 010084068210229221APWX0159411180703, (APWX01594); 1011-9050-000, 010084068210229221APWW0146411170823, (APWW01464); 1011-9050-000, 10084068210229221APWW0162511170927, (APWW01625); 1011-9050-000, 10084068210229221APWW0162611170927, (APWW01626); 1011-9050-000, 10084068210229221APWW0162811170927, (APWW01628); 1011-9050-000, 10084068210229221APWW0208811171206, (APWW02088); 1011-9050-000, 10084068210229221APWW0208911171206, (APWW02089); 1011-9050-000, 10084068210229221APWW0209011171206, (APWW02090); 1011-9050-000, 10084068210229221APWW0209111171212, (APWW02091); 1011-9050-000, 10084068210229221APWW0209211171212, (APWW02092); 1011-9050-000, 10084068210229221APWW0209311171212, (APWW02093); 1011-9050-000, 10084068210229221APWW0209411171212, (APWW02094); 1011-9050-000, 10084068210229221APWW0209511171212, (APWW02095); 1011-9050-000, 10084068210229221APWW0209611171212, (APWW02096); 1011-9050-000, 10084068210229221APWW0209711171212, (APWW02097); 1011-9050-000, 10084068210229221APWW0209811171212, (APWW02098); 1011-9050-000, 10084068210229221APWW0209911171212, (APWW02099); 1011-9050-000, 10084068210229221APWW0210011171212, (APWW02100); 1011-9050-000, 10084068210229221APWW0210111171212, (APWW02101); 1011-9050-000, 10084068210229221APWW0210211171212, (APWW02102); 1011-9050-000, 10084068210229221APWW0210311171212, (APWW02103); 1011-9050-000, 010084068210229221APWX0086711180328, (APWX00867); 1011-9050-000, 010084068210229221APWX0086811180328, (APWX00868); 1011-9050-000, 010084068210229221APWX0086911180328, (APWX00869); 1011-9050-000, 010084068210229221APWX0087011180328, (APWX00870); 1011-9050-000, 010084068210229221APWX0087111180328, (APWX00871); 1011-9050-000, 010084068210229221APWX0087211180328, (APWX00872); 1011-9050-000, 010084068210229221APWX0036911180131, (APWX00369); 1011-9050-000, 010084068210229221APWX0143411180618, (APWX01434). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Upgraded Aisys anesthesia devices and Aisys CS2 anesthesia devices with PSVPro Spontaneous Breathing Modes do not provide the PSVPro ventilation mode feature of transitioning to SIMV PCV backup ventilation mode. PSVPro Spontaneous Breathing Modes do not provide the feature of transitioning to SIMV PCV backup ventilation mode for a patient who stops spontaneous breathing if the PSVPro ventilation mode is in use prior to the clinician starting the Cycling Procedure.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare, LLC
- Distributor
- Worldwide distribution - US nationwide in the states of CA, CO, DC, FL, GA, HI, IA, IL, IN, KY, LA, ,MA, MD, MI, MN, MO, MT, NC, NE, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV. Countries of Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Iceland, India, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea (Republic of), Kuwait, Macedonia, Mexico, Netherlands, New Zealand, Nigeria, Norway, Peru, Portugal, Qatar, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam.
- Lot
- Model Number, UDI Number, (Serial Number): 1011-9050-000, 010084068210229221APWX0132811180608, (APWX01328); 1011-9050-000, 010084068210229221APWX0048411180208, (APWX00484); 1011-9050-000, 010084068210229221APWW0214911171214, (APWW02149); 1011-9050-000, 10084068210229221APWW0215011171214, (APWW02150); 1011-9050-000, 010084068210229221APWX0026111180118, (APWX00261); 1011-9050-000, 010084068210229221APWX0026211180118, (APWX00262); 1011-9050-000, 010084068210229221APWX0140911180618, (APWX01409); 1011-9050-000, 010084068210229221APWX0141011180618, (APWX01410); 1011-9050-000, 010084068210229221APWX0141111180618, (APWX01411); 1011-9050-000, 010084068210229221APWX0141211180618, (APWX01412); 1011-9050-000, 010084068210229221APWX0141311180618, (APWX01413); 1011-9050-000, 010084068210229221APWX0141411180618, (APWX01414); 1011-9050-000, 010084068210229221APWX0141511180618, (APWX01415); 1011-9050-000, 010084068210229221APWX0141611180618, (APWX01416); 1011-9050-000, 010084068210229221APWX0141711180618, (APWX01417); 1011-9050-000, 010084068210229221APWX0141811180618, (APWX01418); 1011-9050-000, 010084068210229221APWX0141911180618, (APWX01419); 1011-9050-000, 010084068210229221APWX0142011180618, (APWX01420); 1011-9050-000, 010084068210229221APWX0142111180618, (APWX01421); 1011-9050-000, 010084068210229221APWX0142211180618, (APWX01422); 1011-9050-000, 010084068210229221APWX0026311180118, (APWX00263); 1011-9050-000, 010084068210229221APWX0026411180118, (APWX00264); 1011-9050-000, 010084068210229221APWX0026511180118, (APWX00265); 1011-9050-000, 010084068210229221APWW0214011171214, (APWW02140); 1011-9050-000, 010084068210229221APWW0215611171214, (APWW02156); 1011-9050-000, 010084068210229221APWW0215711171214, (APWW02157); 1011-9050-000, 010084068210229221APWX0158911180703, (APWX01589); 1011-9050-000, 010084068210229221APWX0159011180703, (APWX01590); 1011-9050-000, 010084068210229221APWX0159111180703, (APWX01591); 1011-9050-000, 010084068210229221APWX0159211180703, (APWX01592); 1011-9050-000, 010084068210229221APWX0159311180703, (APWX01593); 1011-9050-000, 010084068210229221APWX0159411180703, (APWX01594); 1011-9050-000, 010084068210229221APWW0146411170823, (APWW01464); 1011-9050-000, 10084068210229221APWW0162511170927, (APWW01625); 1011-9050-000, 10084068210229221APWW0162611170927, (APWW01626); 1011-9050-000, 10084068210229221APWW0162811170927, (APWW01628); 1011-9050-000, 10084068210229221APWW0208811171206, (APWW02088); 1011-9050-000, 10084068210229221APWW0208911171206, (APWW02089); 1011-9050-000, 10084068210229221APWW0209011171206, (APWW02090); 1011-9050-000, 10084068210229221APWW0209111171212, (APWW02091); 1011-9050-000, 10084068210229221APWW0209211171212, (APWW02092); 1011-9050-000, 10084068210229221APWW0209311171212, (APWW02093); 1011-9050-000, 10084068210229221APWW0209411171212, (APWW02094); 1011-9050-000, 10084068210229221APWW0209511171212, (APWW02095); 1011-9050-000, 10084068210229221APWW0209611171212, (APWW02096); 1011-9050-000, 10084068210229221APWW0209711171212, (APWW02097); 1011-9050-000, 10084068210229221APWW0209811171212, (APWW02098); 1011-9050-000, 10084068210229221APWW0209911171212, (APWW02099); 1011-9050-000, 10084068210229221APWW0210011171212, (APWW02100); 1011-9050-000, 10084068210229221APWW0210111171212, (APWW02101); 1011-9050-000, 10084068210229221APWW0210211171212, (APWW02102); 1011-9050-000, 10084068210229221APWW0210311171212, (APWW02103); 1011-9050-000, 010084068210229221APWX0086711180328, (APWX00867); 1011-9050-000, 010084068210229221APWX0086811180328, (APWX00868); 1011-9050-000, 010084068210229221APWX0086911180328, (APWX00869); 1011-9050-000, 010084068210229221APWX0087011180328, (APWX00870); 1011-9050-000, 010084068210229221APWX0087111180328, (APWX00871); 1011-9050-000, 010084068210229221APWX0087211180328, (APWX00872); 1011-9050-000, 010084068210229221APWX0036911180131, (APWX00369); 1011-9050-000, 010084068210229221APWX0143411180618, (APWX01434);
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACentricity PACS Software Version 7.0 SP0.0.4.7August 3, 2022
- FDASIGNA Premier magnetic resonance scanner, model 5748519.July 27, 2022
- FDAGE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images andJuly 13, 2022
- FDABack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019July 6, 2022
- FDAAisys Anesthesia System. Sold under the following product names: AISYS, AISYS 10.X TO 11.X UPGRADE KIT DUTCH,November 21, 2018
- FDAOviTex 1S Reinforced BioScaffold 4x8cm, Part Number F10256-0408GNovember 21, 2018
Questions about this recall
Was Aisys CS2 Anesthesia System. Sold under the following product names: AISYS, AISYS 10.X TO 11.X UPGRADE KIT DUTCH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH FOR USA, AISYS 10.X TO 11.X UPGRADE KIT FRENCH, AISYS 10.X TO 11.X UPGRADE KIT GERMAN, AISYS 10.X TO 11.X UPGRADE KIT JAPANESE. Product Usage: The GE Datex-Ohmeda Aisys Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). The device is intended for volume or pressure control ventilation. The Aisys is not suitable for use in a MRI environment. recalled?
Yes. FDA published recall Z-0412-2019 on November 21, 2018. Reported hazard: Upgraded Aisys anesthesia devices and Aisys CS2 anesthesia devices with PSVPro Spontaneous Breathing Modes do not provide the PSVPro ventilation mode feature of transitioning to SIMV PCV backup ventilation mode. PSVPro Spontaneous Breathing Modes do not provide the feature of transitioning to SIMV PCV backup ventilation mode for a patient who stops spontaneous breathing if the PSVPro ventilation mode is in use prior to the clinician starting the Cycling Procedure.
What should I do if I have this product?
Voluntary: Firm initiated · Letter