4.5mm PLEO Osteotomy Guide. OrthoPediatrics Corp. markets a pediatric, extension osteotomy bone plate system. This system is marketed under the brand name OrthoPediatrics PediLock Extension Osteotomy Plate ( PLEO ). The product included in this recall is a Class I, 4.5mm PLEO Osteotomy Guide Instrument, common name: Guide, Surgical, Instrument. The PediLoc® Extension Osteotomy Plate is part of the OrthoPediatrics~PediLoc® system. The PediLoc® Extension Osteotomy Plate consists of 3.5mm and 4.5mm plates, left and right side specific. The PediLoc® Extension Osteotomy Plate is designed to fit the anatomy of the distal femur in children and adolescents, above the distal femoral physis. The distal section of the plate is designed so that three locking screws can be placed in the distal bone segment at an angle parallel to the distal femoral physis. The proximal portion of the plate is designed to accept locking or non-locking screws. The subject instrument of this recall is an adjunct instrument for the convenience of some surgeons who choose to use it. This instrument is not illustrated in the Surgical Technique, nor required to perform the procedure.
Published November 28, 2012 · Recall initiated: October 18, 2012
Summary
FDA records medical device Class II recall Z-0432-2013, published November 28, 2012 and initiated October 18, 2012. Recalling firm: OrthoPediatrics Corp; Status: Terminated; Reported quantity: 24; Lot: Part number: 01-1050-0066 and 01-1050-0067 Lot numbers: 7861001 and 7861101 only. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Complaint received from sales representative that the drill guide tubes would not screw into replenishment instrument. Upon investigation, it has been determined that the device was not manufactured to design specifications and will not function as intended.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- OrthoPediatrics Corp
- Distributor
- Nationwide Distribution-including the states of OH, NC, CO, FL, IN, MI, and TX.
- Lot
- Part number: 01-1050-0066 and 01-1050-0067 Lot numbers: 7861001 and 7861101 only
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOrthopedics 6.0MM ROD, COCR, SINGLE HEX, 500MM LENGTH-The RESPONSE" 5.5/6.0 Spine System is intended forAugust 5, 2020
- FDAOrthoPediatrics ACL Reconstruction System ShieldLoc 7mm implant component model number 10-1008-4070 - ProductJuly 15, 2020
- FDAModel No. 01-1503-0032 (Adolescent Nail Attachment Bolt). Component of the Pediatric Nailing Platform. ProductNovember 6, 2019
- FDAResponse BandLoc 5.5/6.0 Tensioner Instrument The Tensioner Instrument (01-1102-1000) is utilized to allow theJune 14, 2017
- FDAStyker Orthobiologics Vita Plasma Separator, a component of Vitagel Surgical Hemostat. Sterile, single useNovember 28, 2012
- FDASenTec V-Sign Membrane Changer (VS-MC), 9 pieces. The SenTec V-Sign Membrane Changer is indicated for use withNovember 28, 2012
Questions about this recall
Was 4.5mm PLEO Osteotomy Guide. OrthoPediatrics Corp. markets a pediatric, extension osteotomy bone plate system. This system is marketed under the brand name OrthoPediatrics PediLock Extension Osteotomy Plate ( PLEO ). The product included in this recall is a Class I, 4.5mm PLEO Osteotomy Guide Instrument, common name: Guide, Surgical, Instrument. The PediLoc® Extension Osteotomy Plate is part of the OrthoPediatrics~PediLoc® system. The PediLoc® Extension Osteotomy Plate consists of 3.5mm and 4.5mm plates, left and right side specific. The PediLoc® Extension Osteotomy Plate is designed to fit the anatomy of the distal femur in children and adolescents, above the distal femoral physis. The distal section of the plate is designed so that three locking screws can be placed in the distal bone segment at an angle parallel to the distal femoral physis. The proximal portion of the plate is designed to accept locking or non-locking screws. The subject instrument of this recall is an adjunct instrument for the convenience of some surgeons who choose to use it. This instrument is not illustrated in the Surgical Technique, nor required to perform the procedure. recalled?
Yes. FDA published recall Z-0432-2013 on November 28, 2012. Reported hazard: Complaint received from sales representative that the drill guide tubes would not screw into replenishment instrument. Upon investigation, it has been determined that the device was not manufactured to design specifications and will not function as intended.
What should I do if I have this product?
Voluntary: Firm initiated · Letter