SenTec V-Sign Membrane Changer Insert (VS-MCI), 5 pieces. The SenTec V-Sign Membrane Changer Insert is indicated for use with the V-Sign Membrane Changer only. The V-sign Sensor is indicated for use with the SenTec Digital Monitor when continuous non-invasive monitoring of carbon dioxide tension, oxygen saturation, and pulse rate are required for adult and pediatric patients.

Published November 28, 2012 · Recall initiated: November 5, 2012

Summary

FDA records medical device Class II recall Z-0434-2013, published November 28, 2012 and initiated November 5, 2012. Recalling firm: Sentec AG; Status: Terminated; Reported quantity: 265 boxes (5 per box); Lot: Lot numbers: 1D0F, 1D58, 1D80, 1DCA, 1DEA, 1E5F. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A manufacturing defect was observed for the membrane changer insert that does not allow electrolyte to be present under the new membrane. The PCO2 performance of a sensor without electrolyte will be deteriorated.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Sentec AG
Distributor
Distributed only in MO.
Lot
Lot numbers: 1D0F, 1D58, 1D80, 1DCA, 1DEA, 1E5F
Risk classification
Class II
Area (FDA)
device
Country
Switzerland

Related recalls

Questions about this recall

Was SenTec V-Sign Membrane Changer Insert (VS-MCI), 5 pieces. The SenTec V-Sign Membrane Changer Insert is indicated for use with the V-Sign Membrane Changer only. The V-sign Sensor is indicated for use with the SenTec Digital Monitor when continuous non-invasive monitoring of carbon dioxide tension, oxygen saturation, and pulse rate are required for adult and pediatric patients. recalled?

Yes. FDA published recall Z-0434-2013 on November 28, 2012. Reported hazard: A manufacturing defect was observed for the membrane changer insert that does not allow electrolyte to be present under the new membrane. The PCO2 performance of a sensor without electrolyte will be deteriorated.

What should I do if I have this product?

Voluntary: Firm initiated · Letter