Back to search
FDAFDA · Food and drugs·Recall number Z-0501-2025

OrganOx metra -Intended as a Normothermic Machine Perfusion system of donor transport livers. Model/Catalog Number: D0003 The OrganOx metra is a transportable device intended to be used to sustain donor livers destined for transplantation in a functioning state for a total preservation time of up to 12 hours.

Published November 27, 2024 · Recall initiated: October 15, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Battery movement led to damage of one of the battery cells causing it to short to the battery retention bracket

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
OrganOx Ltd
Distributor
California, Texas, Ohio, Florida, New York, Massachusetts, Virginia, Oklahoma, DC, South Carolina and Georgia.
Lot
UDI-DI: 05060462240029 Serial numbers 511, 512, 514, 515, 516, 517, 518, 519, 520, 522, 523, 524, 525, 527, 528, 529, 530, 531, 533, 535, 536, 537
Risk classification
Class II
Area (FDA)
device
Country
United Kingdom

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card