FDAFDA · Food and drugs·Recall number Z-0521-2013
Novocastra Ready-to-Use Mouse Monoclonal Antibody Terminal Deoxynuceotidyl Transferase Staining Reagent; an in vitro staining reagent for the qualitative identification by light microscopy of Terminal Deoxynucleotidyl Transferase molecules in paraffin sections; Catalog No. RTU-TdT-339; packaged in 7 mL vials; Leica Biosystems Newcastle Ltd., Balliol Business Park West, Benton Lane, Newcastle Upon Tyne NE12 8EW, United Kingdom. RTU-TdT-339 is intended for the qualitative identification by light microscopy of Terminal Deoxynucleotidyl Transferase molecules in paraffin sections. The clinical interpretation of any staining or its absence should be complemented by morphological studies using proper controls and should be evaluated within the context of the patient's clinical history and other diagnostic tests by a qualified pathologist.
Published December 19, 2012 · Recall initiated: November 14, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
RTU-TdT-339 is not stable up to the expiry date on the product labeling, affecting the staining intensity.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Leica Microsystems, Inc.
- Distributor
- Worldwide Distribution -- USA including Arkansas, California, Florida, Illinois, Maryland, Michigan, Minnesota, Missouri, North Carolina, Oregon, Rhode Island, Tennessee, Texas and Wisconsin, and the country of Canada.
- Lot
- Catalog No. RTU-TdT-339, lot numbers 6001845, expiry 2012-11 and 6009645, expiry 2014-04 The recall also included the folowing expired lots: 804804, 804805, 804806, 804809, 804810, 804812
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States