Novocastra Ready-to-Use Mouse Monoclonal Antibody Terminal Deoxynuceotidyl Transferase Staining Reagent; an in vitro staining reagent for the qualitative identification by light microscopy of Terminal Deoxynucleotidyl Transferase molecules in paraffin sections; Catalog No. RTU-TdT-339; packaged in 7 mL vials; Leica Biosystems Newcastle Ltd., Balliol Business Park West, Benton Lane, Newcastle Upon Tyne NE12 8EW, United Kingdom. RTU-TdT-339 is intended for the qualitative identification by light microscopy of Terminal Deoxynucleotidyl Transferase molecules in paraffin sections. The clinical interpretation of any staining or its absence should be complemented by morphological studies using proper controls and should be evaluated within the context of the patient's clinical history and other diagnostic tests by a qualified pathologist.
Published December 19, 2012 · Recall initiated: November 14, 2012
Summary
FDA records medical device Class II recall Z-0521-2013, published December 19, 2012 and initiated November 14, 2012. Recalling firm: Leica Microsystems, Inc.; Status: Terminated; Reported quantity: 322 vials; Lot: Catalog No. RTU-TdT-339, lot numbers 6001845, expiry 2012-11 and 6009645, expiry 2014-04 The recall also included the folowing expired lots: 804804, 804805, 804806, 804809, 804810, 804812. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
RTU-TdT-339 is not stable up to the expiry date on the product labeling, affecting the staining intensity.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Leica Microsystems, Inc.
- Distributor
- Worldwide Distribution -- USA including Arkansas, California, Florida, Illinois, Maryland, Michigan, Minnesota, Missouri, North Carolina, Oregon, Rhode Island, Tennessee, Texas and Wisconsin, and the country of Canada.
- Lot
- Catalog No. RTU-TdT-339, lot numbers 6001845, expiry 2012-11 and 6009645, expiry 2014-04 The recall also included the folowing expired lots: 804804, 804805, 804806, 804809, 804810, 804812
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABOND Ready-To-Use Primary Antibody CDX2 (EP25), REF PA0375September 22, 2021
- FDABOND Polymer Refine Detection Kit, Catalog No. DS9800. This detection system is for in vitro diagnostic use.December 18, 2019
- FDALeica M530 OHX Microscope System Supplied with 100 - 120 Vac.August 15, 2018
- FDANussloch GmbH ASP6025 Tissue ProcessorJuly 12, 2017
- FDASTERRAD CYCLESURE 24 Biological Indicator (BI) product, P/N 14324. Validation kits with Sterrad Cyclesure 24December 19, 2012
- FDACobas b 123 POC system cobas b 123 POC system is fully automated POC system for whole blood in vitroDecember 19, 2012
Questions about this recall
Was Novocastra Ready-to-Use Mouse Monoclonal Antibody Terminal Deoxynuceotidyl Transferase Staining Reagent; an in vitro staining reagent for the qualitative identification by light microscopy of Terminal Deoxynucleotidyl Transferase molecules in paraffin sections; Catalog No. RTU-TdT-339; packaged in 7 mL vials; Leica Biosystems Newcastle Ltd., Balliol Business Park West, Benton Lane, Newcastle Upon Tyne NE12 8EW, United Kingdom. RTU-TdT-339 is intended for the qualitative identification by light microscopy of Terminal Deoxynucleotidyl Transferase molecules in paraffin sections. The clinical interpretation of any staining or its absence should be complemented by morphological studies using proper controls and should be evaluated within the context of the patient's clinical history and other diagnostic tests by a qualified pathologist. recalled?
Yes. FDA published recall Z-0521-2013 on December 19, 2012. Reported hazard: RTU-TdT-339 is not stable up to the expiry date on the product labeling, affecting the staining intensity.
What should I do if I have this product?
Voluntary: Firm initiated · Letter