FDAFDA · Food and drugs·Recall number Z-0537-2022
The DIXI Medical Microdeep Depth Electrode (Microdeep Depth Electrode or Electrode) is a smooth electrode that is 0.8 mm in diameter with a rounded (hemispherical) tip. The electrode has multiple platinum/iridium contacts that are 2 mm long, with a total exploration length varying from 16 mm to 80.5 mm according to the electrode reference. Microdeep Depth Electrodes are intended for temporary use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain.
Published February 9, 2022 · Recall initiated: December 9, 2021
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential deformation of the internal packaging (blister pack) could compromise sterile barrier of the device when removed from the Tyvek bag
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- DIXI MEDICAL USA
- Distributor
- US Nationwide distribution.
- Lot
- All lots UDI: 03664539000077
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States