FDAFDA · Food and drugs·Recall number Z-0541-2022

The DIXI Medical Microdeep Depth Electrode (Microdeep Depth Electrode or Electrode) is a smooth electrode that is 0.8 mm in diameter with a rounded (hemispherical) tip. The electrode has multiple platinum/iridium contacts that are 2 mm long, with a total exploration length varying from 16 mm to 80.5 mm according to the electrode reference. Microdeep Depth Electrodes are intended for temporary use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain.

Published February 9, 2022 · Recall initiated: December 9, 2021

Summary

FDA records medical device Class II recall Z-0541-2022, published February 9, 2022 and initiated December 9, 2021. Recalling firm: DIXI MEDICAL USA; Status: Terminated; Reported quantity: 1792 units; Lot: All lots UDI:03664539000176. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Potential deformation of the internal packaging (blister pack) could compromise sterile barrier of the device when removed from the Tyvek bag

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
DIXI MEDICAL USA
Distributor
US Nationwide distribution.
Lot
All lots UDI:03664539000176
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was The DIXI Medical Microdeep Depth Electrode (Microdeep Depth Electrode or Electrode) is a smooth electrode that is 0.8 mm in diameter with a rounded (hemispherical) tip. The electrode has multiple platinum/iridium contacts that are 2 mm long, with a total exploration length varying from 16 mm to 80.5 mm according to the electrode reference. Microdeep Depth Electrodes are intended for temporary use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain. recalled?

Yes. FDA published recall Z-0541-2022 on February 9, 2022. Reported hazard: Potential deformation of the internal packaging (blister pack) could compromise sterile barrier of the device when removed from the Tyvek bag

What should I do if I have this product?

Voluntary: Firm initiated · Letter