The DIXI Medical Microdeep Depth Electrode (Microdeep Depth Electrode or Electrode) is a smooth electrode that is 0.8 mm in diameter with a rounded (hemispherical) tip. The electrode has multiple platinum/iridium contacts that are 2 mm long, with a total exploration length varying from 16 mm to 80.5 mm according to the electrode reference. Microdeep Depth Electrodes are intended for temporary use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain.
Published February 9, 2022 · Recall initiated: December 9, 2021
Summary
FDA records medical device Class II recall Z-0541-2022, published February 9, 2022 and initiated December 9, 2021. Recalling firm: DIXI MEDICAL USA; Status: Terminated; Reported quantity: 1792 units; Lot: All lots UDI:03664539000176. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential deformation of the internal packaging (blister pack) could compromise sterile barrier of the device when removed from the Tyvek bag
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- DIXI MEDICAL USA
- Distributor
- US Nationwide distribution.
- Lot
- All lots UDI:03664539000176
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDA3-Component Cold Tabs (Acetaminophen 325 mg, Guaifenesin 200 mg, Phenylephrine HCl 5mg) Bulk Container,February 16, 2022
- FDAThe DIXI Medical Microdeep Depth Electrode (Microdeep Depth Electrode or Electrode) is a smooth electrodeFebruary 9, 2022
- FDAKWIK-STIK, Enterobacter cloacae subsp. cloacae derived from ATCC 13047, REF 0323P. Packaged 2 units per pack.February 9, 2022
- FDAVagus Nerve Stimulation (VNS) Therapy Leads, PerenniaDURA, Model: 303. Model 303-20 PereniaDURA Lead (2.0 mmFebruary 9, 2022
- FDAThe EnPlace is a single use trans-vaginal pelvic floor repair system which enables delivery of a tissue anchorFebruary 9, 2022
- FDAWayfare Dairy Free Butter Salted & Whipped,February 9, 2022
Questions about this recall
Was The DIXI Medical Microdeep Depth Electrode (Microdeep Depth Electrode or Electrode) is a smooth electrode that is 0.8 mm in diameter with a rounded (hemispherical) tip. The electrode has multiple platinum/iridium contacts that are 2 mm long, with a total exploration length varying from 16 mm to 80.5 mm according to the electrode reference. Microdeep Depth Electrodes are intended for temporary use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain. recalled?
Yes. FDA published recall Z-0541-2022 on February 9, 2022. Reported hazard: Potential deformation of the internal packaging (blister pack) could compromise sterile barrier of the device when removed from the Tyvek bag
What should I do if I have this product?
Voluntary: Firm initiated · Letter