SOL SYS L 9IN CALC 2.25/18 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foam sleeve for added protection. The assembly is then placed and sealed in an inner and outer tray for aseptic transfer. The sealed package is placed into a folding carton, labeled, and shrink wrapped. Intended Use: SOLUTION System Calcar Stems are used in total hip arthroplasty (THA) to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components.
Published December 26, 2012 · Recall initiated: December 3, 2012
Summary
FDA records medical device Class II recall Z-0582-2013, published December 26, 2012 and initiated December 3, 2012. Recalling firm: DePuy Orthopaedics, Inc.; Status: Terminated; Reported quantity: 1732 total all product codes; Lot: Product Code: 157225180 and Lot codes: B19HN1000, B19HNA000, B3FAF1000, BF6A21000, C32GEA, C5PRY1000, C61CD1000, CD1EW1000, EE5K61000, X33GC1000, X49FV1000, X4KF91000, X4KGA1000, X4KGB1000, X5SB41000, X5SB4A000, X5SB4B000, X5SB4C000, Y14GJ1000, Y14GJA000, YD3H31000, YD3HY1000, YG8B11000, YH9E21000, and YH9E2B000. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
An Engineering Analysis has determined the Solution Calcar Stem has the potential to result in impingement with a variety of heads. This impingement may not allow the stem and head to lock as intended. The amount of possible impingement varies depending on the femoral head. Worst case condition: Calcar stem with -2mm offset articuleze maximum of 0.937mm radial impingement. It is evident that a
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- DePuy Orthopaedics, Inc.
- Distributor
- Nationwide Distribution including AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WY and GUAYNABO, PR
- Lot
- Product Code: 157225180 and Lot codes: B19HN1000, B19HNA000, B3FAF1000, BF6A21000, C32GEA, C5PRY1000, C61CD1000, CD1EW1000, EE5K61000, X33GC1000, X49FV1000, X4KF91000, X4KGA1000, X4KGB1000, X5SB41000, X5SB4A000, X5SB4B000, X5SB4C000, Y14GJ1000, Y14GJA000, YD3H31000, YD3HY1000, YG8B11000, YH9E21000, and YH9E2B000.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAATTUNE AFFIXIUM Cementless Fixed Bearing Knee with 3DP Technology, Size 6. Component of the ATTUNE Total KneeApril 10, 2024
- FDAAttune Revision Limb Preservation System (LPS) Tibial Insert XSM 12MM, Part Number 151760212March 29, 2023
- FDAAttune Revision Limb Preservation System (LPS) Tibial Insert SM 12MM, Part Number 151760312March 29, 2023
- FDAATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the bone cuts in preparationJuly 27, 2022
- FDASOL SYS L 9IN CALC 2.25/16.5 Packaging: Product is packed within a poly protector, with protection added asDecember 26, 2012
- FDASOL SYS L 9IN CALC 2.25/19.5 Packaging: Product is packed within a poly protector, with protection added asDecember 26, 2012
Questions about this recall
Was SOL SYS L 9IN CALC 2.25/18 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foam sleeve for added protection. The assembly is then placed and sealed in an inner and outer tray for aseptic transfer. The sealed package is placed into a folding carton, labeled, and shrink wrapped. Intended Use: SOLUTION System Calcar Stems are used in total hip arthroplasty (THA) to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. recalled?
Yes. FDA published recall Z-0582-2013 on December 26, 2012. Reported hazard: An Engineering Analysis has determined the Solution Calcar Stem has the potential to result in impingement with a variety of heads. This impingement may not allow the stem and head to lock as intended. The amount of possible impingement varies depending on the femoral head. Worst case condition: Calcar stem with -2mm offset articuleze maximum of 0.937mm radial impingement. It is evident that a
What should I do if I have this product?
Voluntary: Firm initiated · Letter