FlexLab, Accelerator a3600 and Aptio Automation are modular systems designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.

Published November 23, 2016 · Recall initiated: October 4, 2016

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Inpeco received several notifications from the field about burnt schuko sockets. The customers reported that some smoke was seen coming from the damaged sockets.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Inpeco S.A.
Distributor
Worldwide Distribution - US including Texas and New York. Foreign Consignees: Germany, Italy, France, Austria and Spain.
Lot
Shucko Socket with P/N: 693822160, Model: CMSA/HETTICH - STORAGE - i2000SR, Length: 3 meters; Shucko Socket with P/N: 693822170, Model: CMSA/HETTICH - STORAGE - i2000SR, Length: 5 meters; Shucko Socket with P/N: 693822190, Model: CMSA/HETTICH - STORAGE - i2000SR, Length: 7 meters; Shucko Socket with P/N: 0C00005640, Model: CMSA/HETTICH - STORAGE - i2000SR, Length: 2 meters. The Serial number of the impacted products globally distributed are the following: ACP.009, ACP.024, ACP.039, ACP.054, ACP.069, ACP.084, ACP.100, ACP.115 , ACP.010, ACP.025, ACP.040, ACP.055, ACP.070, ACP.085, ACP.101, ACP.116, ACP.011, ACP.026, ACP.041, ACP.056, ACP.071, ACP.086, ACP.102, ACP.117, ACP.012, ACP.027, ACP.042, ACP.057, ACP.072,ACP.087, ACP.103, ACP.118 , ACP.013, ACP.028, ACP.043, ACP.058, ACP.073, ACP.088, ACP.104, ACP.119, ACP.014, ACP.029, ACP.044, ACP.059, ACP.074, ACP.089, ACP.105, ACP.120, ACP.015, ACP.030, ACP.045, ACP.060, ACP.075, ACP.090, ACP.110, ACP.121, ACP.016, ACP.031, ACP.046, ACP.061, ACP.076, ACP.091, ACP.107, ACP.122 ,ACP.017, ACP.032, ACP.047, ACP.062, ACP.077, ACP.092, ACP.108, ACP.123 ,ACP.018, ACP.033, ACP.048, ACP.063, ACP.078, ACP.093, ACP.109, ACP.124, ACP.019, ACP.034, ACP.049, ACP.064, ACP.079, ACP.094, ACP.110, ACP.125, ACP.020, ACP.035, ACP.050, ACP.065, ACP.080, ACP.096, ACP.111, ACP.126 ,ACP.021, ACP.036, ACP.051, ACP.066, ACP.081, ACP.097, ACP.112, ACP.127 , ACP.022, ACP.037, ACP.052, ACP.067, ACP.082, ACP.098, ACP.113, ACP.128, ACP.023, ACP.038, ACP.053, ACP.068, ACP.083, ACP.099, ACP.114, ACP.129, FLX.103, FLX.0120, FLX.0131, FLX.0137, FLX.0143, FLX.0161, FLX.0166, FLX.0179, FLX.0104, FLX.0124, FLX.0134, FLX.0140, FLX.0152, FLX.0163, FLX.0175, FLX.0108, FLX.0125, FLX.0135, FLX.0141 FLX.0155, FLX.0164, FLX.0176, FLX.0115, FLX.0126, FLX.0136, FLX.0142, FLX.0158, FLX.0165, FLX.0178, AP2.0001, AP2.0004, AP2.0006, AP2.0008, AP2.0010, AP2.0012, AP2.0014, AP2.0003, AP2.0002, AP2.0005, AP2.0007, AP2.0009, AP2.0011, AP2.0013, AP2.0015, FLX.0008, FLX.0133, FLX.0149, FLX.0153, FLX.0159, FLX.0167, FLX.0170, FLX.0174, FLX.0069, FLX.0139, FLX.0150 FLX.0156, FLX.0162, FLX.0168, FLX.0172, FLX.0179, FLX.0132, FLX.0147, FLX.0151, FLX.0157, FLX.0166, FLX.0169, FLX.0173 ,FLX.0121, FLX.0138, FLX.0144, FLX.0145, FLX.0146, FLX.0148, FLX.0176 and FLX.0181.
Risk classification
Class II
Area (FDA)
device
Country
Switzerland