Inpeco FlexLab, ARCHITECT cSystem c8000/c16000 Interface Module, Part Numbers: FLX-208-00, FLX-208-10 - Product Usage: The FlexLab (aka Accelerator a3600) is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.

Published May 20, 2020 · Recall initiated: April 7, 2020

Summary

FDA records medical device Class II recall Z-2003-2020, published May 20, 2020 and initiated April 7, 2020. Recalling firm: Inpeco S.A.; Status: Terminated; Reported quantity: 82; Lot: Serial Numbers: AML.001, FLX.0104, FLX.0105, FLX.0106, FLX.0107, FLX.0108, FLX.0112, FLX.0114, FLX.0115, FLX.0116, FLX.0118, FLX.0119, FLX.0120, FLX.0124, FLX.0125, FLX.0126, FLX.0131, FLX.0133, FLX.0134, FLX.0135, FLX.0136, FLX.0140, FLX.0141, FLX.0142, FLX.0143, FLX.015, FLX.0152, FLX.0155, FLX.0158, FLX.0161, FLX.0163, FLX.0164, FLX.017, FLX.0175, FLX.0178, FLX.018, FLX.0184, FLX.0189, FLX.019, FLX.0196, FLX.020, FLX.021, FLX.023, FLX.026, FLX.027, FLX.030, FLX.032, FLX.033, FLX.038, FLX.040, FLX.044, FLX.045, FLX.047, FLX.051, FLX.052, FLX.053, FLX.054, FLX.055, FLX.058, FLX.059, FLX.063, FLX.064, FLX.065, FLX.066, FLX.067, FLX.068, FLX.071, FLX.072, FLX.074, FLX.075, FLX.076, FLX.077, FLX.079, FLX.080, FLX.081, FLX.091, FLX.092, FLX.095, FLX.099, FLX.101, FLX.102, FLX.103. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

When analyzer sample carousel lid remains open or manual process is too long, Module timeout expiry errors could occur causing cross contamination of samples.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Inpeco S.A.
Distributor
Worldwide distribution - US Nationwide including in the states of Texas, West Virginia, South Carolina, Delaware, Louisiana, Illinois, Georgia, Florida, New York, Virginia, Arkansas, Wisconsin, California, Minnesota, Missouri, North Carolina, Utah, Michigan, Ohio, Alabama, Oklahoma, Tennessee, Kansas, Hawaii, Arizona, Washington
Lot
Serial Numbers: AML.001, FLX.0104, FLX.0105, FLX.0106, FLX.0107, FLX.0108, FLX.0112, FLX.0114, FLX.0115, FLX.0116, FLX.0118, FLX.0119, FLX.0120, FLX.0124, FLX.0125, FLX.0126, FLX.0131, FLX.0133, FLX.0134, FLX.0135, FLX.0136, FLX.0140, FLX.0141, FLX.0142, FLX.0143, FLX.015, FLX.0152, FLX.0155, FLX.0158, FLX.0161, FLX.0163, FLX.0164, FLX.017, FLX.0175, FLX.0178, FLX.018, FLX.0184, FLX.0189, FLX.019, FLX.0196, FLX.020, FLX.021, FLX.023, FLX.026, FLX.027, FLX.030, FLX.032, FLX.033, FLX.038, FLX.040, FLX.044, FLX.045, FLX.047, FLX.051, FLX.052, FLX.053, FLX.054, FLX.055, FLX.058, FLX.059, FLX.063, FLX.064, FLX.065, FLX.066, FLX.067, FLX.068, FLX.071, FLX.072, FLX.074, FLX.075, FLX.076, FLX.077, FLX.079, FLX.080, FLX.081, FLX.091, FLX.092, FLX.095, FLX.099, FLX.101, FLX.102, FLX.103
Risk classification
Class II
Area (FDA)
device
Country
Switzerland

Related recalls

Questions about this recall

Was Inpeco FlexLab, ARCHITECT cSystem c8000/c16000 Interface Module, Part Numbers: FLX-208-00, FLX-208-10 - Product Usage: The FlexLab (aka Accelerator a3600) is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation. recalled?

Yes. FDA published recall Z-2003-2020 on May 20, 2020. Reported hazard: When analyzer sample carousel lid remains open or manual process is too long, Module timeout expiry errors could occur causing cross contamination of samples.

What should I do if I have this product?

Voluntary: Firm initiated · Letter