Landauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-000
Published January 3, 2024 · Recall initiated: November 15, 2023
Summary
FDA records medical device Class II recall Z-0616-2024, published January 3, 2024 and initiated November 15, 2023. Recalling firm: Landauer; Status: Completed; Reported quantity: 584 units; Lot: All serial numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Completed — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
MicroSTARii readers may be contributing to the nanodot dosimeter measurement inaccuracy in two possible ways. The first is related to the interaction of the reader s LED beam profile with the nanoDot dosimeter. The second is debris formation due to drawer actuation, resulting in a change of reader response that may not be detectable unless known dose QCs are being run regularly.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Landauer
- Distributor
- US and Australia, Brazil, Canada, China, France, Germany, India, Japan, Mexico, Rep of Korea, Taiwan, Thailand, Viet Nam
- Lot
- All serial numbers.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAmicroStar Reader, Model Numbers: 10015-000 (marketed in EMEA as 100, 102, 110, and 201), 10016-000 (marketedMarch 27, 2024
- FDAMedtronic SynchroMed II, Model 8637-40, Programmable pumpJanuary 3, 2024
- FDAHookup Accessories used with MEDIVATORS ADVANTAGE PLUS and ADVANTAGE PLUS Pass Thru Automated EndoscopeJanuary 3, 2024
- FDAnanoDot D2DNS, for custom calibrate, OTO, Model Numbers 03034-OTO; radiation monitoring dosimeter used withSeptember 20, 2023
- FDAnanoDot kit (contains nanoDot A, B, C), Model Number 03500-000; radiation monitoring dosimeter used with theSeptember 20, 2023
- FDAnanoDot D2DNS, for custom calibrate, OTO, Model Number 03033-OTO; radiation monitoring dosimeter used with theSeptember 20, 2023
Questions about this recall
Was Landauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-000 recalled?
Yes. FDA published recall Z-0616-2024 on January 3, 2024. Reported hazard: MicroSTARii readers may be contributing to the nanodot dosimeter measurement inaccuracy in two possible ways. The first is related to the interaction of the reader s LED beam profile with the nanoDot dosimeter. The second is debris formation due to drawer actuation, resulting in a change of reader response that may not be detectable unless known dose QCs are being run regularly.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail