nanoDot kit (contains nanoDot A, B, C), Model Number 03500-000; radiation monitoring dosimeter used with the microSTAR readers
Published September 20, 2023 · Recall initiated: July 12, 2023
Summary
FDA records medical device Class II recall Z-2553-2023, published September 20, 2023 and initiated July 12, 2023. Recalling firm: Landauer; Status: Completed; Reported quantity: 16760 units; Lot: UDI/DI 0860003399903, all batch numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Completed — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Landauer
- Distributor
- Worldwide
- Lot
- UDI/DI 0860003399903, all batch numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAmicroStar Reader, Model Numbers: 10015-000 (marketed in EMEA as 100, 102, 110, and 201), 10016-000 (marketedMarch 27, 2024
- FDALandauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-000January 3, 2024
- FDATRP Natural Eyes, Aging Eye Relief, Sterile Eye Drops, Homeopathic 0.33 FL OZ (10 mL), Manufactured in Jordan,September 27, 2023
- FDATRP Natural Eyes, Allergy Eyes Relief, Sterile Eye Drops, Homeopathic 0.33 FL OZ (10 mL), Manufactured inSeptember 27, 2023
- FDAnanoDot D2DNS, for custom calibrate, OTO, Model Numbers 03034-OTO; radiation monitoring dosimeter used withSeptember 20, 2023
- FDAnanoDot D2DNS, for custom calibrate, OTO, Model Number 03033-OTO; radiation monitoring dosimeter used with theSeptember 20, 2023
Questions about this recall
Was nanoDot kit (contains nanoDot A, B, C), Model Number 03500-000; radiation monitoring dosimeter used with the microSTAR readers recalled?
Yes. FDA published recall Z-2553-2023 on September 20, 2023. Reported hazard: LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
What should I do if I have this product?
Voluntary: Firm initiated · Letter