Back to search
FDAFDA · Food and drugs·Recall number Z-2538-2023

nanoDot D2DNS, for custom calibrate, OTO, Model Number 03033-OTO; radiation monitoring dosimeter used with the microSTAR readers

Published September 20, 2023 · Recall initiated: July 12, 2023

Status when published: Completed — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Landauer
Distributor
Worldwide
Lot
UDI/DI 0860003399903, all batch numbers
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?