Nova StatStrip/StatSensor Hospital Blood Glucose Meter; Nova StatStrip Lactate Hospital Meter System; Nova StatSensor Creatinine Hospital Meter, Rechargeable Lithium Batteries Single: 46626; 4 Pack: 46827; 5 Pack: 50436. For Point-Of-Care usage in the quantitative determination of Glucose in whole blood.
Published January 15, 2014 · Recall initiated: October 10, 2013
Summary
FDA records medical device Class II recall Z-0650-2014, published January 15, 2014 and initiated October 10, 2013. Recalling firm: Nova Biomedical Corporation; Status: Terminated; Reported quantity: est. 200,000; Lot: All serial numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Revised Battery Instructions for Removal and Expiration Date due to battery swelling and cell leakage
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Nova Biomedical Corporation
- Distributor
- Nationwide Distribution.
- Lot
- All serial numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAStatStrip Glucose Hospital Meter System. Model No: 66739. Glucose test system.September 23, 2026
- FDAStatStrip Glucose and B- Ketone meter. Model No: 63910. Glucose test system.September 23, 2026
- FDAStatStrip Glucose Hospital Meter System. Model No: 63685. Glucose test system.September 23, 2026
- FDAStatStrip Glucose and B- Ketone meter. Model No: 63683. Glucose test system.September 23, 2026
- FDACatalys Precision Laser System (United States) - Catalys-U: Catalys Precision Laser System (International) -January 15, 2014
- FDAThe Vantage Digital Panoramic System delivers high-value features standard on every system: - FiveJanuary 15, 2014
Questions about this recall
Was Nova StatStrip/StatSensor Hospital Blood Glucose Meter; Nova StatStrip Lactate Hospital Meter System; Nova StatSensor Creatinine Hospital Meter, Rechargeable Lithium Batteries Single: 46626; 4 Pack: 46827; 5 Pack: 50436. For Point-Of-Care usage in the quantitative determination of Glucose in whole blood. recalled?
Yes. FDA published recall Z-0650-2014 on January 15, 2014. Reported hazard: Revised Battery Instructions for Removal and Expiration Date due to battery swelling and cell leakage
What should I do if I have this product?
Voluntary: Firm initiated · Letter