(l) Visum Halogen Surgical Light, (2) Visum LED Surgical Light, (3) Stryker Flat Panel and Navigation Arm The Spring Arm Circlip is part of the Stryker Spring Arm used in multiple configurations as a component of the (1)Visum Halogen Surgical Light, (2)Visum LED Surgical Light, and (3)Stryker Flat Panel and Navigation Arm. The Spring Arm that is subject to this field action suspends loads, is a conduit for power/data, and facilitates multi-directional movement. The Spring Arms that have been identified within scope are replacement parts that have potentially been serviced incorrectly.
Published January 16, 2013 · Recall initiated: September 25, 2012
Summary
FDA records medical device Class II recall Z-0657-2013, published January 16, 2013 and initiated September 25, 2012. Recalling firm: Stryker Communications, Inc.; Status: Terminated; Reported quantity: United States: 1,082 units and Outside United States: 199 units; Lot: Code Number and Product Name: 0682000112-SPRING ARM, LIGHT 12-18KG, 0682000211 - FLAT PANEL SPRING ARM, ONDAL SUSPENSION, 15-21KG, 0682000212 - FLAT PANEL SPRING ARM, ONDAL SUSP., 12-18KG, 0682000226 - SPRING ARM,3.5-7KG,NAVIGATION, 0682000228 - SPRING ARM, LIGHT 9-15KG, 0682001062 - PKG EDS LIGHT SUSPENSION CENTRAL AXIS 5P/FP, 0682001270 - 9-POLE SPRING ARM, AC 2000,15-21 KG, 0682001274 - 5-POLE SPRING ARM AC 2000 15-21 KG, 0682400034 - FLAT PANEL SPRING ARM, EDS SUSPENSION, 15-21KG, 0682400059 - UNIVERSAL YOKE SPRING ARM RETROFIT KIT, 06824000591 - UNIVERSAL YOKE & SPRING ARM RETROFIT KIT, INT, 0682400151 - LOW CEILING SPRING ARM, 5P SIDE MOUNT, 0682400152 - LOW CEILING SPRING ARM, 9P SIDE MOUNT, 0682400169 - 5-POLE SPRING ARM AC 2000 9-15KG, CARDINAL, 0682400207 - STANDARD SPRING ARM WITH BEARING, 5P, 0682400208 - STANDARD SPRING ARM WITH BEARING, 9P, and 0682400365 - FLAT PANEL SPRING ARM WITH ERA ADAPTER BEARING. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Stryker has become aware that there is a low likelihood of incomplete engagement of the Spring Arm Circlip during service or replacement of the Spring Arm. If the Spring Arm Circlip is not fully seated, the Spring Arm could separate and fall along with the surgical lights or monitors mounted to the end of the Spring Arm. To date, there have been no serious injuries reported from the incomplete e
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Stryker Communications, Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) and Internationally to Australia, Japan, Taiwan, Canada, United Kingdom, France, Netherlands, Spain, Mexico, India, Korea, Singapore, New Zealand, Hong Kong, Philippines and Brazil.
- Lot
- Code Number and Product Name: 0682000112-SPRING ARM, LIGHT 12-18KG, 0682000211 - FLAT PANEL SPRING ARM, ONDAL SUSPENSION, 15-21KG, 0682000212 - FLAT PANEL SPRING ARM, ONDAL SUSP., 12-18KG, 0682000226 - SPRING ARM,3.5-7KG,NAVIGATION, 0682000228 - SPRING ARM, LIGHT 9-15KG, 0682001062 - PKG EDS LIGHT SUSPENSION CENTRAL AXIS 5P/FP, 0682001270 - 9-POLE SPRING ARM, AC 2000,15-21 KG, 0682001274 - 5-POLE SPRING ARM AC 2000 15-21 KG, 0682400034 - FLAT PANEL SPRING ARM, EDS SUSPENSION, 15-21KG, 0682400059 - UNIVERSAL YOKE SPRING ARM RETROFIT KIT, 06824000591 - UNIVERSAL YOKE & SPRING ARM RETROFIT KIT, INT, 0682400151 - LOW CEILING SPRING ARM, 5P SIDE MOUNT, 0682400152 - LOW CEILING SPRING ARM, 9P SIDE MOUNT, 0682400169 - 5-POLE SPRING ARM AC 2000 9-15KG, CARDINAL, 0682400207 - STANDARD SPRING ARM WITH BEARING, 5P, 0682400208 - STANDARD SPRING ARM WITH BEARING, 9P, and 0682400365 - FLAT PANEL SPRING ARM WITH ERA ADAPTER BEARING.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAccolade TMZF Plus 127 degree Neck Angle V40 Hip Stem #2.5 and Accolade TMZF Plus 127 degree Neck Angle V40January 16, 2013
- FDAAutoPulse Resuscitation System Model 100, Mfg by:Zoll Circulation Sunnyvale, CA. The AutoPulse Model 100January 16, 2013
- FDAFlowGate Balloon Guide Catheter; Manufactured by Concentric Medical, Mountain View, CA.January 16, 2013
- FDAThe TM Ardis Interbody System Inserter Stainless Steel, Rubber, Silicone Non Sterile Rx only. The TM ArdisJanuary 16, 2013
- FDAUltraFill DBM labeled in part: TISSUENET Sterilized by T10^6 Description: UltraFill DBM Size: 10cc ProductJanuary 16, 2013
- FDAPT113950 PT HYBRID Glenoid Post Regenerex porous titanium construct Sterile R Prosthesis, Shoulder,January 16, 2013
Questions about this recall
Was (l) Visum Halogen Surgical Light, (2) Visum LED Surgical Light, (3) Stryker Flat Panel and Navigation Arm The Spring Arm Circlip is part of the Stryker Spring Arm used in multiple configurations as a component of the (1)Visum Halogen Surgical Light, (2)Visum LED Surgical Light, and (3)Stryker Flat Panel and Navigation Arm. The Spring Arm that is subject to this field action suspends loads, is a conduit for power/data, and facilitates multi-directional movement. The Spring Arms that have been identified within scope are replacement parts that have potentially been serviced incorrectly. recalled?
Yes. FDA published recall Z-0657-2013 on January 16, 2013. Reported hazard: Stryker has become aware that there is a low likelihood of incomplete engagement of the Spring Arm Circlip during service or replacement of the Spring Arm. If the Spring Arm Circlip is not fully seated, the Spring Arm could separate and fall along with the surgical lights or monitors mounted to the end of the Spring Arm. To date, there have been no serious injuries reported from the incomplete e
What should I do if I have this product?
Voluntary: Firm initiated · Letter