Biopsy Needle FPU Kit, Product # 9731754, UDI: 00643169341517
Published December 18, 2019 · Recall initiated: October 30, 2019
Summary
FDA records medical device Class II recall Z-0658-2020, published December 18, 2019 and initiated October 30, 2019. Recalling firm: Medtronic Navigation, Inc.; Status: Ongoing; Reported quantity: 12 disposable biopsy needles; Lot: Lot Numbers: 0009665221 0009709452 0009960538. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm has become aware that due to a manufacturing issue impacting the biopsy needle depth stop, there exist the potential that some biopsy needle depth stops, when tightened, may not securely tighten to the biopsy needle. This could result in a delay in surgery.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medtronic Navigation, Inc.
- Distributor
- US: Alabama, Alaska,Arizona,Arkansas,California,Colorado,Connecticut,Delaware,District of Columbia,Florida,Georga, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine,Maryland,Massachusetts,Michigan,Minnesota,Mississippi,Missouri,Montana,Nebraska,Nevada,New Hampshire,New Jersey,New Mexico,New York,North Carolina,North Dakota,Ohio,Oklahoma,Oregon,Pennsylvania,South Carolina,South Dakota,Tennessee,Texas,Utah,Vermont,Virginia,Washington,West Virginia,Wisconsin,Wyoming, OUS: None: Australia, Austria,Bangladesh,Belgium,Bosnia And Herzegovina,Brazil,Bulgaria,Canada,Canary Islands,China,Colombia,Costa Rica,Croatia,Czech Republic,Denmark,Ecuador,Egypt,Finland,France,Germany,Greece,Hungary,Iceland,India,Indonesia,Ireland,Israel,Italy,Japan,Jordan,Kazakhstan,Kenya,Korea, Republic Of,Latvia,Lebanon,Luxembourg,Malaysia,Martinique,Mexico,Netherlands,New Zealand,Pakistan,Philippines,Poland,Portugal,Puerto Rico,Qatar,Reunion,Romania,Russian Federation,Saudi Arabia,Serbia,Slovakia,Slovenia,South Africa,Spain,Sweden,Switzerland,Thailand,Turkey,United Arab Emirates,United Kingdom
- Lot
- Lot Numbers: 0009665221 0009709452 0009960538
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMedtronic Biopsy Needle Kit, REF: 9733068, Passive, Sterile EO, RxonlyOctober 28, 2026
- FDAMedtronic NDI Passive Spheres REF: 8801075, Sterile EO, 5/Tray 12 Pk., RxonlyOctober 28, 2026
- FDAMedtronic NDI Passive Spheres REF: 8801071, Sterile EO, 1/Tray 12 pk., RxonlyOctober 28, 2026
- FDAMedtronic NDI Passive Spheres REF: 8801074, Sterile EO, 4/Tray 12 Pk., RxonlyOctober 28, 2026
- FDALigamax 5 mm Endoscopic Multiple Clip ApplierDecember 18, 2019
- FDAOral-B, Practitioner Series, 5% Sodium Fluoride Varnish, 500 units (40mL/each) - Product Usage: For dentinalDecember 18, 2019
Questions about this recall
Was Biopsy Needle FPU Kit, Product # 9731754, UDI: 00643169341517 recalled?
Yes. FDA published recall Z-0658-2020 on December 18, 2019. Reported hazard: The firm has become aware that due to a manufacturing issue impacting the biopsy needle depth stop, there exist the potential that some biopsy needle depth stops, when tightened, may not securely tighten to the biopsy needle. This could result in a delay in surgery.
What should I do if I have this product?
Voluntary: Firm initiated · Letter