LINK STEM with Microporous Surface MP Reconstruction Prosthesis, CEMENTLESS (Size XS-0, D: 12mm, L: 160mm), Reference Number 172-916/12 Product Usage: The LINK® STEM with Microporous Surface MP® Reconstruction Prosthesis, CEMENTLESS System is indicated for patients with mobility-limiting diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures. The LINK® STEM with Microporous Surface MP® Reconstruction Prosthesis, CEMENTLESS System is indicated for the following conditions: Revision arthroplasty due to juxta-articular bone defects Revision of loosened femoral prosthesis components involving extensive bone resorption of the proximal femur and widening of the medullary cavity or marked thinning of proximal femoral cortical bone Revision of loosened femoral prosthesis components with periprosthetic/subprosthetic fracture Deformed proximal femur due to fractures or osteotomies Correction of bone deficiencies, e.g. due to tumors Large post-revision and post-trauma segmental bone defects Oncological and revision surgery from tibial to hip area (in conjunction with Endo-Model® SL Rotational and Hinge Knee Prostheses) The device is intended for cemented and cementless use.

Published January 23, 2019 · Recall initiated: December 7, 2018

Summary

FDA records medical device Class II recall Z-0733-2019, published January 23, 2019 and initiated December 7, 2018. Recalling firm: Waldemar Link GmbH & Co. KG (Mfg Site); Status: Terminated; Reported quantity: 11; Lot: Lot numbers: 1435077 1515029 1718103 1435077 150401/0217 1734168 1726056 1718103 1734168 1515029. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The reaming guide and the guide rod can't be screwed into the intended depth. The reason for this is a reduced depth of the core hole in the stem. A reaming guide that is not screwed in completely may lead to a proximal preparation in the depth that is not sufficient and to a neck that can't be positioned correctly. The correct fit of the neck can't be verified by means of the guide rod.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Waldemar Link GmbH & Co. KG (Mfg Site)
Distributor
US Nationwide in the states of: FL, GA, IA, IN, KS, MI, MN, and TN.
Lot
Lot numbers: 1435077 1515029 1718103 1435077 150401/0217 1734168 1726056 1718103 1734168 1515029
Risk classification
Class II
Area (FDA)
device
Country
Germany

Related recalls

Questions about this recall

Was LINK STEM with Microporous Surface MP Reconstruction Prosthesis, CEMENTLESS (Size XS-0, D: 12mm, L: 160mm), Reference Number 172-916/12 Product Usage: The LINK® STEM with Microporous Surface MP® Reconstruction Prosthesis, CEMENTLESS System is indicated for patients with mobility-limiting diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures. The LINK® STEM with Microporous Surface MP® Reconstruction Prosthesis, CEMENTLESS System is indicated for the following conditions: Revision arthroplasty due to juxta-articular bone defects Revision of loosened femoral prosthesis components involving extensive bone resorption of the proximal femur and widening of the medullary cavity or marked thinning of proximal femoral cortical bone Revision of loosened femoral prosthesis components with periprosthetic/subprosthetic fracture Deformed proximal femur due to fractures or osteotomies Correction of bone deficiencies, e.g. due to tumors Large post-revision and post-trauma segmental bone defects Oncological and revision surgery from tibial to hip area (in conjunction with Endo-Model® SL Rotational and Hinge Knee Prostheses) The device is intended for cemented and cementless use. recalled?

Yes. FDA published recall Z-0733-2019 on January 23, 2019. Reported hazard: The reaming guide and the guide rod can't be screwed into the intended depth. The reason for this is a reduced depth of the core hole in the stem. A reaming guide that is not screwed in completely may lead to a proximal preparation in the depth that is not sufficient and to a neck that can't be positioned correctly. The correct fit of the neck can't be verified by means of the guide rod.