Back to search
FDAFDA · Food and drugs·Recall number Z-0752-2024

Calibration Serum Level 3 CAL2351

Published January 24, 2024 · Recall initiated: November 20, 2023

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

CK Total in Calibration Serum Level 3, CAL2351, is running with a positive bias on RX Series instruments compared to other methods experience a shift in Quality Control and patient sample recovery of up to 13%.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
Randox Laboratories Ltd.
Distributor
US Nationwide and Puerto Rico.
Lot
GTIN: 05055273200966 Lot Numbers: 1214UE, 1249UE, 1260UE, 1262UE, 1268UE, 1297UE, 1298UE, 1315UE
Risk classification
Class II
Area (FDA)
device
Country
Ireland