Calibration Serum Level 3 CAL2351
Published January 24, 2024 · Recall initiated: November 20, 2023
Summary
FDA records medical device Class II recall Z-0752-2024, published January 24, 2024 and initiated November 20, 2023. Recalling firm: Randox Laboratories Ltd.; Status: Ongoing; Reported quantity: 269 kits; Lot: GTIN: 05055273200966 Lot Numbers: 1214UE, 1249UE, 1260UE, 1262UE, 1268UE, 1297UE, 1298UE, 1315UE. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
CK Total in Calibration Serum Level 3, CAL2351, is running with a positive bias on RX Series instruments compared to other methods experience a shift in Quality Control and patient sample recovery of up to 13%.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Randox Laboratories Ltd.
- Distributor
- US Nationwide and Puerto Rico.
- Lot
- GTIN: 05055273200966 Lot Numbers: 1214UE, 1249UE, 1260UE, 1262UE, 1268UE, 1297UE, 1298UE, 1315UE
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Ireland
Related recalls
- FDARX Series CYSC- Cystatin C Reagent -Intended for in vitro diagnostic (IVD) use as Randox Cystatin C assay onJuly 31, 2024
- FDAClinical Chemistry Calibration Serum (Catalog number Cal2351) that is 20x5ml lyophilized human based serumMay 15, 2024
- FDAEvidence MultiSTAT with software v 3.7-Analyzer intended for the qualitative determination of parent drugMay 15, 2024
- FDAMicroalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay Catalog Number: MA1567April 24, 2024
- FDACalibration Serum Level 3. CAL 3, REF CAL2351. Human Assayed Multi-Sera Level 2 (HN1530) Human AssayedJanuary 24, 2024
- FDACalibration Serum Level 3 - IVD For use as a Calibrator in clinical chemistry assays Catalog Number: CAL2351January 24, 2024
Questions about this recall
Was Calibration Serum Level 3 CAL2351 recalled?
Yes. FDA published recall Z-0752-2024 on January 24, 2024. Reported hazard: CK Total in Calibration Serum Level 3, CAL2351, is running with a positive bias on RX Series instruments compared to other methods experience a shift in Quality Control and patient sample recovery of up to 13%.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail