Calibration Serum Level 3 - IVD For use as a Calibrator in clinical chemistry assays Catalog Number: CAL2351
Published January 24, 2024 · Recall initiated: November 20, 2023
Summary
FDA records medical device Class II recall Z-0753-2024, published January 24, 2024 and initiated November 20, 2023. Recalling firm: Randox Laboratories Ltd.; Status: Ongoing; Reported quantity: 108 kits; Lot: GTIN: 05055273200966 Lot/Batch No: 1214UE Exp. Date: 28 Nov 2023. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Target for Alkaline Phosphatase (ALP) for the AMP optimized to IFCC 37®C method has been mis-assigned for the RX Series instruments by approximately 10% on Quality Control and patient samples
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Randox Laboratories Ltd.
- Distributor
- US Nationwide including Puerto Rico.
- Lot
- GTIN: 05055273200966 Lot/Batch No: 1214UE Exp. Date: 28 Nov 2023
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Ireland
Related recalls
- FDARX Series CYSC- Cystatin C Reagent -Intended for in vitro diagnostic (IVD) use as Randox Cystatin C assay onJuly 31, 2024
- FDAClinical Chemistry Calibration Serum (Catalog number Cal2351) that is 20x5ml lyophilized human based serumMay 15, 2024
- FDAEvidence MultiSTAT with software v 3.7-Analyzer intended for the qualitative determination of parent drugMay 15, 2024
- FDAMicroalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay Catalog Number: MA1567April 24, 2024
- FDACalibration Serum Level 3 CAL2351January 24, 2024
- FDACardinalHealth Suction Canister Liner, MEDI-VAC CRD, REF 65651-510January 24, 2024
Questions about this recall
Was Calibration Serum Level 3 - IVD For use as a Calibrator in clinical chemistry assays Catalog Number: CAL2351 recalled?
Yes. FDA published recall Z-0753-2024 on January 24, 2024. Reported hazard: Target for Alkaline Phosphatase (ALP) for the AMP optimized to IFCC 37®C method has been mis-assigned for the RX Series instruments by approximately 10% on Quality Control and patient samples
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail