FDAFDA · Food and drugs·Recall number Z-0753-2024

Calibration Serum Level 3 - IVD For use as a Calibrator in clinical chemistry assays Catalog Number: CAL2351

Published January 24, 2024 · Recall initiated: November 20, 2023

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Target for Alkaline Phosphatase (ALP) for the AMP optimized to IFCC 37¿C method has been mis-assigned for the RX Series instruments by approximately 10% on Quality Control and patient samples

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
Randox Laboratories Ltd.
Distributor
US Nationwide including Puerto Rico.
Lot
GTIN: 05055273200966 Lot/Batch No: 1214UE Exp. Date: 28 Nov 2023
Risk classification
Class II
Area (FDA)
device
Country
Ireland

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card