GE Healthcare Optima XR220amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-12; 2) 5555000-5; 3) 5555000-6; 4) A0659PB; 5) to be provided
Published January 15, 2025 · Recall initiated: November 25, 2024
Summary
FDA records medical device Class II recall Z-0812-2025, published January 15, 2025 and initiated November 25, 2024. Recalling firm: GE Medical Systems, LLC; Status: Ongoing; Reported quantity: 281 units; Lot: 1) 5555000-12: UDI/DI none, Serial Numbers: 082407100532; UDI/DI to be provided, Serial Numbers: 082407160512; 2) 5555000-5: UDI/DI 840682115261, Serial Numbers: 1557XR0205, 352867XRA, 478765CDP1, 509962KV220, 601358OPTX220, 620421P3, 813872OPT3, SB2201700099WK; UDI/DI none, Serial Numbers: 479314220, 00028MOB12, 203694SB3, 205820DP1, 208666KDM, 209334OPT1, 210703M3, 240566PX2, 240964OPT1, 248465DP3, 256HHPORT2, 256HHPORT3, 256MHPORT1, 281MSL220, 281MSL2202, 305KEN220A, 306766DMOB2, 310680PORTDR2, 314251NAV7, 325574OPTITBD2, 330489DIGPORT, 333OPTI2, 333OPTI3, 333OPTI4, 352596OPT1, 352LEEDP1, 352VILDP1, 386FHODP3, 401OPTI1, 405636P2, 405636P240, 405752OPT1, 405752OPT2, 405752OPT4, 405DELCITY220, 407303OPTIMAX220, 408259AP6, 417347DPORT1, 417556P1, 417556P2, 417556P3, 419251DP1, 419251DP2, 419383DP1, 434654OPTIMA, 435CVXR220, 480MAYO220A, 501364OPTI1, 501364OPTI2, 509474DR1, 510437OPT1, 510727OPT3, 512SAMCSB, 513870DP1, 520TMCOPT2, 541426XR220, 559447OPT1, 559784OPT2, 561694OPT1, 561694OPT2, 561965DPORT5, 561965DPORT7, 573882DP1, 573882DP2, 573882DP3, 580421OPT1, 601288SB1, 601288SB2, 601288SB3, 601288SB6, 601928D220, 608364OPMX220, 613938XR3, 626457GEPORT1, 703369MDIG2, 703664M7, 706596SFP2, 706651DP, 715838DP2, 718206OPT240C, 718267DP5, 718670PTDR1, 720UCHSB7, 727848SB1, 731541SB1, 731541SB1A, 731541SB2, 731541SB3, 732324PA1, 732828DP11, 757388DP4, 770385220D1, 772468OPT1, 785776AMX220P1, 804281SMDM1, 806296OPTI220, 813302OPT3, 814837OPT, 815300M220C, 817927JPSXR220G, 817MCAXR220B, 828213DXP1, 828262P2202, 830RHHM1, 830RHHM2, 843847220B, 850325DP2, 863699OP220, 903886XR220, 905527XR220, 905845MOB4, 908685DP2, 909464220P1, 910615OPTIMA1, 910615OPTIMA2, 915521OPTIMA1, 915595OPT1, 940898XR220A, 940898XR240D, 985867M1, AXXXXX38, ORMCPORTA, ORMCPORTB, ORMCPORTC, ORMCPORTD, TAMXDRP03, WD0005, XR1036454WK5; 3) 5555000-6: UDI/DI 840682115261, Serial Numbers: 06051MOB06, 228809XR220, 228818XR220, 270798XR220B, 281KINGOPT3, 303450SB3, 303695SB3, 416323MOB3, 530888OPT1, 540981ODP3R, 540981ODP9, 615769OPTX220, 713529MDR220, 717782OPT4, 7982XR0192, 802775SB1, 813681OPT2, 918307OPT2, MHSWOPT2201, SA2513MM03, SB2201600071WK, SB2201700085WK, SJRO220P1, SJRO220P3, UTAHJAZZ, XR1864CO; UDI/DI none, Serial Numbers: 083007001584712, 0850070865, 0850070866, 0850070870, 0850070913, 0856070254, 0856070267, 0856070268, 091026DEMO2245, 10181MOB16, 140024RX14, 206598USB7, 209943OTP1, 214PH220G, 228575D220, 254HHSMOB1, 256429PORT2, 270688SB5, 281338BAPOR1, 281338CL221, 281338MDR220, 281348KWDM2, 281348KWDM3, 281KINGOPT1, 281MWB2202, 281MWB2203, 304255OPT220, 304623OPT220, 308865GSDP1, 309779RDIG2, 314525NAV15, 314525NAV16, 318675PW1, 318675PW2, 318769RDPT1, 318769RDPT2, 330480DP1, 330480DP2, 330480DP3, 330480DP4, 370OPTI1, 409938DR1220, 412232DMX2, 412372DMX1, 412647DMX2, 419291DP4, 419893DP1, 423493DP2, 432685OPT2, 434799GE220, 469764XR220H, 504454XR221, 504785MD1, 504873TMD8, 513732DP1, 563742NDIG1, 574267XR220, 586263DP1, 586493DP1, 601798D220, 614257DP4, 615284STMDR2, 615284STMDR3, 615316DP1, 615342DP1, 615391SUR, 702SIEXR221, 702SIEXR226, 713359BSDM3, 713440DP1, 713441OPCDX1, 714966GEPORT2, 720UCHSB9, 740592OPTIMA2, 77522MOB14, 801479SB2, 810342DP2, 817255NHXR220A, 817HUG220C, 817PLAZA220A, 817SAGERXR220A, 865541DP2, 865545DP1, 865545DP3, 903CSM222, 904OPMCDP3, 914737PORTWK3, 915577XR220, 916UCDAMXDR23, 916UCDAMXDR27, 918579OPT2, 919774DIGPRT, 928757XR222, 936539DM1, 936539DM2, 949517DR1, 949764DR2, 949764DR3, 949764DR4, 972420XR220B, 972420XR220C, 972566XR220F, 979MGXR220, LLPM220P1, MHKOPT220, SA1056MM06, SA2161MM04, XR1031657WK8, ZA1067RX30; 4) A0659PB: UDI/DI 840682115261, Serial Numbers: 702880EROP1; UDI/DI none, Serial Numbers: 206598USB6; 5) to be provided: UDI/DI none, Serial Numbers: 240826OPT1, 772SRDP1, 817HUG220B 860545SB2A; UDI/DI to be provided, Serial Numbers: 907852OPT22019, SA1075MM04. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Medical Systems, LLC
- Distributor
- Worldwide
- Lot
- 1) 5555000-12: UDI/DI none, Serial Numbers: 082407100532; UDI/DI to be provided, Serial Numbers: 082407160512; 2) 5555000-5: UDI/DI 840682115261, Serial Numbers: 1557XR0205, 352867XRA, 478765CDP1, 509962KV220, 601358OPTX220, 620421P3, 813872OPT3, SB2201700099WK; UDI/DI none, Serial Numbers: 479314220, 00028MOB12, 203694SB3, 205820DP1, 208666KDM, 209334OPT1, 210703M3, 240566PX2, 240964OPT1, 248465DP3, 256HHPORT2, 256HHPORT3, 256MHPORT1, 281MSL220, 281MSL2202, 305KEN220A, 306766DMOB2, 310680PORTDR2, 314251NAV7, 325574OPTITBD2, 330489DIGPORT, 333OPTI2, 333OPTI3, 333OPTI4, 352596OPT1, 352LEEDP1, 352VILDP1, 386FHODP3, 401OPTI1, 405636P2, 405636P240, 405752OPT1, 405752OPT2, 405752OPT4, 405DELCITY220, 407303OPTIMAX220, 408259AP6, 417347DPORT1, 417556P1, 417556P2, 417556P3, 419251DP1, 419251DP2, 419383DP1, 434654OPTIMA, 435CVXR220, 480MAYO220A, 501364OPTI1, 501364OPTI2, 509474DR1, 510437OPT1, 510727OPT3, 512SAMCSB, 513870DP1, 520TMCOPT2, 541426XR220, 559447OPT1, 559784OPT2, 561694OPT1, 561694OPT2, 561965DPORT5, 561965DPORT7, 573882DP1, 573882DP2, 573882DP3, 580421OPT1, 601288SB1, 601288SB2, 601288SB3, 601288SB6, 601928D220, 608364OPMX220, 613938XR3, 626457GEPORT1, 703369MDIG2, 703664M7, 706596SFP2, 706651DP, 715838DP2, 718206OPT240C, 718267DP5, 718670PTDR1, 720UCHSB7, 727848SB1, 731541SB1, 731541SB1A, 731541SB2, 731541SB3, 732324PA1, 732828DP11, 757388DP4, 770385220D1, 772468OPT1, 785776AMX220P1, 804281SMDM1, 806296OPTI220, 813302OPT3, 814837OPT, 815300M220C, 817927JPSXR220G, 817MCAXR220B, 828213DXP1, 828262P2202, 830RHHM1, 830RHHM2, 843847220B, 850325DP2, 863699OP220, 903886XR220, 905527XR220, 905845MOB4, 908685DP2, 909464220P1, 910615OPTIMA1, 910615OPTIMA2, 915521OPTIMA1, 915595OPT1, 940898XR220A, 940898XR240D, 985867M1, AXXXXX38, ORMCPORTA, ORMCPORTB, ORMCPORTC, ORMCPORTD, TAMXDRP03, WD0005, XR1036454WK5; 3) 5555000-6: UDI/DI 840682115261, Serial Numbers: 06051MOB06, 228809XR220, 228818XR220, 270798XR220B, 281KINGOPT3, 303450SB3, 303695SB3, 416323MOB3, 530888OPT1, 540981ODP3R, 540981ODP9, 615769OPTX220, 713529MDR220, 717782OPT4, 7982XR0192, 802775SB1, 813681OPT2, 918307OPT2, MHSWOPT2201, SA2513MM03, SB2201600071WK, SB2201700085WK, SJRO220P1, SJRO220P3, UTAHJAZZ, XR1864CO; UDI/DI none, Serial Numbers: 083007001584712, 0850070865, 0850070866, 0850070870, 0850070913, 0856070254, 0856070267, 0856070268, 091026DEMO2245, 10181MOB16, 140024RX14, 206598USB7, 209943OTP1, 214PH220G, 228575D220, 254HHSMOB1, 256429PORT2, 270688SB5, 281338BAPOR1, 281338CL221, 281338MDR220, 281348KWDM2, 281348KWDM3, 281KINGOPT1, 281MWB2202, 281MWB2203, 304255OPT220, 304623OPT220, 308865GSDP1, 309779RDIG2, 314525NAV15, 314525NAV16, 318675PW1, 318675PW2, 318769RDPT1, 318769RDPT2, 330480DP1, 330480DP2, 330480DP3, 330480DP4, 370OPTI1, 409938DR1220, 412232DMX2, 412372DMX1, 412647DMX2, 419291DP4, 419893DP1, 423493DP2, 432685OPT2, 434799GE220, 469764XR220H, 504454XR221, 504785MD1, 504873TMD8, 513732DP1, 563742NDIG1, 574267XR220, 586263DP1, 586493DP1, 601798D220, 614257DP4, 615284STMDR2, 615284STMDR3, 615316DP1, 615342DP1, 615391SUR, 702SIEXR221, 702SIEXR226, 713359BSDM3, 713440DP1, 713441OPCDX1, 714966GEPORT2, 720UCHSB9, 740592OPTIMA2, 77522MOB14, 801479SB2, 810342DP2, 817255NHXR220A, 817HUG220C, 817PLAZA220A, 817SAGERXR220A, 865541DP2, 865545DP1, 865545DP3, 903CSM222, 904OPMCDP3, 914737PORTWK3, 915577XR220, 916UCDAMXDR23, 916UCDAMXDR27, 918579OPT2, 919774DIGPRT, 928757XR222, 936539DM1, 936539DM2, 949517DR1, 949764DR2, 949764DR3, 949764DR4, 972420XR220B, 972420XR220C, 972566XR220F, 979MGXR220, LLPM220P1, MHKOPT220, SA1056MM06, SA2161MM04, XR1031657WK8, ZA1067RX30; 4) A0659PB: UDI/DI 840682115261, Serial Numbers: 702880EROP1; UDI/DI none, Serial Numbers: 206598USB6; 5) to be provided: UDI/DI none, Serial Numbers: 240826OPT1, 772SRDP1, 817HUG220B 860545SB2A; UDI/DI to be provided, Serial Numbers: 907852OPT22019, SA1075MM04.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASenographe Pristina 3DNovember 11, 2026
- FDASenographe PristinaNovember 11, 2026
- FDASenographe PristinaNovember 11, 2026
- FDAOptima XR240amx, Mobile Digital Radiographic SystemJuly 22, 2026
- FDAXhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system withJanuary 15, 2025
- FDATec 820 ISO Anesthesia Vaporizer, Reference Number 1177-9820-ISOJanuary 15, 2025
Questions about this recall
Was GE Healthcare Optima XR220amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-12; 2) 5555000-5; 3) 5555000-6; 4) A0659PB; 5) to be provided recalled?
Yes. FDA published recall Z-0812-2025 on January 15, 2025. Reported hazard: GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.
What should I do if I have this product?
Voluntary: Firm initiated · Letter