FDAFDA · Food and drugs·Recall number Z-0842-2014

API Listeria (REF 10300) API Listeria is a standardized system for the identification of Listeria which uses miniaturized tests, as well as a database. The ZYM B reagent is including in API Listeria kit and it is used as an additional reagent for revealing the result of DIM miniaturized biochemical test included in the API Listeria strips. ZYM B reagent is added in the DIM test before reading miniaturized biochemical test including in the API Listeria strips.

Published February 5, 2014 · Recall initiated: November 8, 2013

Summary

FDA records medical device Class II recall Z-0842-2014, published February 5, 2014 and initiated November 8, 2013. Recalling firm: Biomerieux France Chemin De L'Or; Status: Terminated; Reported quantity: 39439 kits (includes ZYM B reagents, API Listeria * API NH kits); Lot: Lot numbers: 1002210010, 1002210011, 1002287930, 1002287931, 1002350760, 1002390940, 1002390941, 1002518070, 1002518071, 1002518072, 1001915090, 1001915091, 1001915092, 1001932340, 1001998120 and 1002129350. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

bioM®rieux identified a visual defect and activity issue on the ZYM B reagent (REF 70493); leading to a false negative results of some API biochemical tests using the ZYM B reagent.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Biomerieux France Chemin De L'Or
Distributor
Worldwide Distribution - USA (nationwide) and Internationally to U.S. Nationwide; Algerie, Argentina, Armenia, Australia, Austria, Azerbaidjian, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bolivia, Brazil, Brunei, Bulgaria, Burkina Faso, Cambodia, Cameroon, Canada, Chad, Chile, China, Colombia, Democratic Republic of Congo, Costa Rica, Cote d'Ivorie, Germany, Czech Republic, Dominican Republic, Dutch Antilles, Dutch Antilles, Ecuador, Egypt, Republica De El Salvador, Estonia, Germany, Ethiopia, France, Gabon, Georgia, Germany, Greece, Guatemala, Guinea, Herzegovina, Honduras, Hungary, India, Germany, Indonesia, Iraq, Israel, Italy, Jamaica, Japan, Iraq, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malaysia, Malaysia, Mali, Malta, Mexico, Romania, Mongolia, Morocco, Mozambique, Singapore, Netherlands, Managua, Niamey Niger, Nigeria, Nigeria, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Uganda, Kingdom of Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sudan, Sweden, Switzerland, Syria, Tanzania, Italy, Thailand, Trinidad & Tobago, Tunisia, Turkey, Uganda, Ukraine, UAE, United Kingdom, Uruguay, Venezuela, Vietnam, Vietnam, Yemen and Zimbabwe
Lot
Lot numbers: 1002210010, 1002210011, 1002287930, 1002287931, 1002350760, 1002390940, 1002390941, 1002518070, 1002518071, 1002518072, 1001915090, 1001915091, 1001915092, 1001932340, 1001998120 and 1002129350.
Risk classification
Class II
Area (FDA)
device
Country
France

Related recalls

Questions about this recall

Was API Listeria (REF 10300) API Listeria is a standardized system for the identification of Listeria which uses miniaturized tests, as well as a database. The ZYM B reagent is including in API Listeria kit and it is used as an additional reagent for revealing the result of DIM miniaturized biochemical test included in the API Listeria strips. ZYM B reagent is added in the DIM test before reading miniaturized biochemical test including in the API Listeria strips. recalled?

Yes. FDA published recall Z-0842-2014 on February 5, 2014. Reported hazard: bioM®rieux identified a visual defect and activity issue on the ZYM B reagent (REF 70493); leading to a false negative results of some API biochemical tests using the ZYM B reagent.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail