Ad-Tech Medical Instrument Corporation Macro Micro Subdural Electrode. For temporary (<30 days) use with recording, monitoring and stimulation equipment, for the recording, monitoring and stimulation of electrical signals on the surface level of the brain. The recording of electrical activity supports definition of the location of epileptogenic foci and brain mapping.
Published March 13, 2013 · Recall initiated: December 18, 2012
Summary
FDA records medical device Class I recall Z-0892-2013, published March 13, 2013 and initiated December 18, 2012. Recalling firm: Ad-Tech Medical Instrument Corporation; Status: Terminated; Reported quantity: 144 (103 US; 41 OUS); Lot: REF numbers (catalog numbers): EG10A-SP10N-000, EG34B-SP10X-000, EGD7A-MP12N-000, EGE4A-SP10N-000, EGE4A-SP10X-000, QG10B-SP10X-000, QG10C-SP10N-000, QG10D-SP10N-000, QG11A-SP10X-000, QG16A-SP10X-000, QG17A-SP10X-000, QG18A-SP10N-000, QG18A-SP10X-000, QG20B-SP10N-000, QG20B-SP10X-000, QG20C-SP10N-000, QG20C-SP10X-000, QG20D-SP10X-200, QG20E-SP10X-000, QG20F-SP10X-000, QG24A-SP10N-000, QG26A-SP10N-000, QG32A-SP10X-000, QG32B-SP10X-000, QG36A-SP10N-000, QG36A-SP10X-000, QG36B-SP10N-000, QG44A-SP10X-000, QG53A-SP10X-000, QG66A-SP85N-000, QG69A-SP10N-000, QG88A-SP10X-000, QG96A-SP10N-000, QG96A-SP10X-000, QG96C-SP10X-000, QG96D-SP10X-000, QGC6A-SP07N-000, QGC8A-SP10N-000, QGC8A-SP10X-000, QGC8B-SP10X-000, QGC8C-SP10X-000, QGC8D-SP10N-000, QGC8E-SP10X-000, QGC8F-SP10X-000, QS20A-SP10N-000, QS40B-SP10X-000, Lot numbers: 208140314, 208140343, 208140344, 208140353, 208140354, 208140362, 208140369, 208140374, 208140381, 208140382, 208140387, 208140390, 208140395, 208140398, 208140399, 208140414, 208140417, 208140421, 208140431, 208140443, 208140445, 208140450, 208140452, 208140453, 208140458, 019020, 208140356R, 208140357R, 208140452R, C010180001, C702201001, C703160006, C704040001, C800250002, C800350004, C801240003, C802220003, C802220005, C802220006, C805210003, C805210004, C908221002, G012111001, G012140002, G015020002, G016240001, G017250004, G017250005, G017250006, G018011001, G018011002, G018011003, G019021002, G019021003, G019150003, G112270004, G114160001, G114160002, G115080002, G115170004, G115280005, G119020004, G211220001, G608060002, G704150001, G805270001, G900211002, G903011001, G903080003, S903080004, S905280003. Recall exanded on 5/2/2013 to include additional Macro Micro Subdural Electrodes: REF numbers (catalog numbers): CG-64BPX-F436, CMMS-22PX-F478, , CG-32BPX-F412, CGIE-98BPX-F341, CGIE-82BPN-F377, CMMS-69PN-F480, CGIE-98BPN-F341, CG-9SN-F496, CG-32BPX-F426, CG-16BPX-F445, CG-16BPX-F446, CG-256PX-F425, CGIE-39BPX-F442, CG-46BPX-F465, CG-64BPX-F419, CG-32BPX-F418. Lot numbers: 20814044, 208140447, 208140463, 208140463, 208140476, 208140463, 08140463, 208140424, 208140439, G119221001,G212040002, G216040002, 218190, 208140431, 208140450, 208140432, 208140433, 208140420, 208140437, 208140450, 208140422, 208140422, 208140422, 208140422. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The microelectrodes are not flush with the silastic substrate, may contain sharp edges capable of abrading the cortical surface, and may have remaining pieces which may fracture and remain in the cortical tissue with the electrode is explanted. Recall expanded on 5/2/2013 to include additional Macro Micro Subdural Electrodes.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ad-Tech Medical Instrument Corporation
- Distributor
- Worldwide Distribution - USA Nationwide including the states of CA, GA, LA, MA, MD, MI, MN, NH, PA, RI, TX, UT, WI, NY IA, WI and the countries of: Australia, CANADA, ENGLAND, JAPAN, KOREA, SPAIN, FRANCE, and SWITZERLAND.
- Lot
- REF numbers (catalog numbers): EG10A-SP10N-000, EG34B-SP10X-000, EGD7A-MP12N-000, EGE4A-SP10N-000, EGE4A-SP10X-000, QG10B-SP10X-000, QG10C-SP10N-000, QG10D-SP10N-000, QG11A-SP10X-000, QG16A-SP10X-000, QG17A-SP10X-000, QG18A-SP10N-000, QG18A-SP10X-000, QG20B-SP10N-000, QG20B-SP10X-000, QG20C-SP10N-000, QG20C-SP10X-000, QG20D-SP10X-200, QG20E-SP10X-000, QG20F-SP10X-000, QG24A-SP10N-000, QG26A-SP10N-000, QG32A-SP10X-000, QG32B-SP10X-000, QG36A-SP10N-000, QG36A-SP10X-000, QG36B-SP10N-000, QG44A-SP10X-000, QG53A-SP10X-000, QG66A-SP85N-000, QG69A-SP10N-000, QG88A-SP10X-000, QG96A-SP10N-000, QG96A-SP10X-000, QG96C-SP10X-000, QG96D-SP10X-000, QGC6A-SP07N-000, QGC8A-SP10N-000, QGC8A-SP10X-000, QGC8B-SP10X-000, QGC8C-SP10X-000, QGC8D-SP10N-000, QGC8E-SP10X-000, QGC8F-SP10X-000, QS20A-SP10N-000, QS40B-SP10X-000, Lot numbers: 208140314, 208140343, 208140344, 208140353, 208140354, 208140362, 208140369, 208140374, 208140381, 208140382, 208140387, 208140390, 208140395, 208140398, 208140399, 208140414, 208140417, 208140421, 208140431, 208140443, 208140445, 208140450, 208140452, 208140453, 208140458, 019020, 208140356R, 208140357R, 208140452R, C010180001, C702201001, C703160006, C704040001, C800250002, C800350004, C801240003, C802220003, C802220005, C802220006, C805210003, C805210004, C908221002, G012111001, G012140002, G015020002, G016240001, G017250004, G017250005, G017250006, G018011001, G018011002, G018011003, G019021002, G019021003, G019150003, G112270004, G114160001, G114160002, G115080002, G115170004, G115280005, G119020004, G211220001, G608060002, G704150001, G805270001, G900211002, G903011001, G903080003, S903080004, S905280003. Recall exanded on 5/2/2013 to include additional Macro Micro Subdural Electrodes: REF numbers (catalog numbers): CG-64BPX-F436, CMMS-22PX-F478, , CG-32BPX-F412, CGIE-98BPX-F341, CGIE-82BPN-F377, CMMS-69PN-F480, CGIE-98BPN-F341, CG-9SN-F496, CG-32BPX-F426, CG-16BPX-F445, CG-16BPX-F446, CG-256PX-F425, CGIE-39BPX-F442, CG-46BPX-F465, CG-64BPX-F419, CG-32BPX-F418. Lot numbers: 20814044, 208140447, 208140463, 208140463, 208140476, 208140463, 08140463, 208140424, 208140439, G119221001,G212040002, G216040002, 218190, 208140431, 208140450, 208140432, 208140433, 208140420, 208140437, 208140450, 208140422, 208140422, 208140422, 208140422
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAD-TECH Spencer Probe Depth Electrode, REF SD06R-AP58X-000 6-Contact 1.12mm (A style) Depth ElectrodeOctober 30, 2024
- FDAAD-TECH Spencer Probe Depth Electrode, REF SD08R-AP58X-000 8-Contact 1.12mm (A style) Depth ElectrodeOctober 30, 2024
- FDAAD-TECH Spencer Probe Depth Electrode, REF SD04R-AP58X-000 4-Contact 1.12mm (A style) Depth ElectrodeOctober 30, 2024
- FDAAD-TECH Drill Sleeve Guides, used with Ad-Tech Placement Kits: (a) REF DSG-90-2.3N (b) REF DSG-6.3-090-2.4NSeptember 25, 2019
- FDAHERBALIFE Formula 1 Instant Healthy Meal, Nutritional Shake Mix ( Vanilla Dream) Packets; net WT 1.83 oz/52 g;March 13, 2013
- FDAPhilips Digital Diagnost Stationary radiographic systemMarch 13, 2013
Questions about this recall
Was Ad-Tech Medical Instrument Corporation Macro Micro Subdural Electrode. For temporary (<30 days) use with recording, monitoring and stimulation equipment, for the recording, monitoring and stimulation of electrical signals on the surface level of the brain. The recording of electrical activity supports definition of the location of epileptogenic foci and brain mapping. recalled?
Yes. FDA published recall Z-0892-2013 on March 13, 2013. Reported hazard: The microelectrodes are not flush with the silastic substrate, may contain sharp edges capable of abrading the cortical surface, and may have remaining pieces which may fracture and remain in the cortical tissue with the electrode is explanted. Recall expanded on 5/2/2013 to include additional Macro Micro Subdural Electrodes.
What should I do if I have this product?
Voluntary: Firm initiated · Letter