Back to search
FDAFDA · Food and drugs·Recall number Z-0895-2020

CS Elite Fastpack, 225 mL, 150 Res Autotransfusion Device - Product Usage: its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell product to either a bag for gravity reinfusion into the patient or to the arterial line of an extracorporeal circuit for reinfusion into the patient.

Published February 5, 2020 · Recall initiated: October 21, 2019

Summary

FDA records medical device Class II recall Z-0895-2020, published February 5, 2020 and initiated October 21, 2019. Recalling firm: Haemonetics Corporation; Status: Terminated; Reported quantity: 42739; Lot: Catalog number/FG Item Number: CSE-FP-225V 1. Specific Lots with Bowl Cracks (7636 Distributed) Lot number: 1902009HTT 1903001HTT 1903004HTT 1804008HTT 1804010HTT 1805004HTT 1805005HTT 1812012HTT 1901004HTT 1901005HTT 1901006HTT 1808007HTT 1808009HTT 1808012HTT 1809001HTT 1809002HTT 1809006HTT 1809009HTT 1810001HTT 1810003HTT 1901005HTT 1902009HTT 1810006HTT 2. Missing instructions to check for cracks in IFU All products/lost manufactured between October 2014 and October 2019. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Potential to develop leaks at the inner core of Cell Saver® 5/5+ and Cell Saver® Elite®/Elite® +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Haemonetics Corporation
Distributor
Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
Lot
Catalog number/FG Item Number: CSE-FP-225V 1. Specific Lots with Bowl Cracks (7636 Distributed) Lot number: 1902009HTT 1903001HTT 1903004HTT 1804008HTT 1804010HTT 1805004HTT 1805005HTT 1812012HTT 1901004HTT 1901005HTT 1901006HTT 1808007HTT 1808009HTT 1808012HTT 1809001HTT 1809002HTT 1809006HTT 1809009HTT 1810001HTT 1810003HTT 1901005HTT 1902009HTT 1810006HTT 2. Missing instructions to check for cracks in IFU All products/lost manufactured between October 2014 and October 2019.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was CS Elite Fastpack, 225 mL, 150 Res Autotransfusion Device - Product Usage: its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell product to either a bag for gravity reinfusion into the patient or to the arterial line of an extracorporeal circuit for reinfusion into the patient. recalled?

Yes. FDA published recall Z-0895-2020 on February 5, 2020. Reported hazard: Potential to develop leaks at the inner core of Cell Saver® 5/5+ and Cell Saver® Elite®/Elite® +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.

What should I do if I have this product?

Voluntary: Firm initiated · Letter