CS5/5+ Fastpacks, 125 mL, 20® Res Autotransfusion Device- Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.
Published February 5, 2020 · Recall initiated: October 21, 2019
Summary
FDA records medical device Class II recall Z-0889-2020, published February 5, 2020 and initiated October 21, 2019. Recalling firm: Haemonetics Corporation; Status: Terminated; Reported quantity: 1340; Lot: Catalog number/FG Item Number: 0265F-00 1. No specific lots with bowl cracks. 2. Missing instructions to check for cracks in IFU All products/lost manufactured between October 2014 and October 2019. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential to develop leaks at the inner core of Cell Saver® 5/5+ and Cell Saver® Elite®/Elite® +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Haemonetics Corporation
- Distributor
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
- Lot
- Catalog number/FG Item Number: 0265F-00 1. No specific lots with bowl cracks. 2. Missing instructions to check for cracks in IFU All products/lost manufactured between October 2014 and October 2019.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACS5/5+ Fastpacks, 225 mL, 20® Res Autotransfusion Device - Product Usage: its related accessory components areFebruary 5, 2020
- FDACS Elite Fastpack, 225 mL, 150 Res Autotransfusion Device - Product Usage: its related accessory componentsFebruary 5, 2020
- FDAJapan DISP SET, 263J, 205J, HAR - Product Usage: its related accessory components are intended to be used forFebruary 5, 2020
- FDAJAPAN CELL SAVER ELITE SET - 225ML - Product Usage: its related accessory components are intended for use toFebruary 5, 2020
- FDA2.4MM TI VA LOCKING SCREW STARDRIVE 18MM STERILE, Part Number 04.210.118TS - Product Usage:Sterile TubeFebruary 5, 2020
- FDA2.4MM TI VA LOCKING SCREW STARDRIVE 10MM STERILE, Part Number 04.210.110TS - Product Usage:Sterile TubeFebruary 5, 2020
Questions about this recall
Was CS5/5+ Fastpacks, 125 mL, 20® Res Autotransfusion Device- Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma. recalled?
Yes. FDA published recall Z-0889-2020 on February 5, 2020. Reported hazard: Potential to develop leaks at the inner core of Cell Saver® 5/5+ and Cell Saver® Elite®/Elite® +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
What should I do if I have this product?
Voluntary: Firm initiated · Letter