Japan DISP SET, 263J, 205J, HAR - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma

Published February 5, 2020 · Recall initiated: October 21, 2019

Summary

FDA records medical device Class II recall Z-0906-2020, published February 5, 2020 and initiated October 21, 2019. Recalling firm: Haemonetics Corporation; Status: Terminated; Reported quantity: 1584; Lot: Catalog number/FG Item Number: 263J-SS-SET 1. Specific Lots with Bowl Cracks (110 Distributed) Lot number: 0718027-0119029-08170710119003; 0718027-0818086-08170711018045; 0718027-1118031-08170711118019; 0718027-1118081-08170710119084 2. Missing instructions to check for cracks in IFU All products/lost manufactured between October 2014 and October 2019. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Potential to develop leaks at the inner core of Cell Saver® 5/5+ and Cell Saver® Elite®/Elite® +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Haemonetics Corporation
Distributor
Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
Lot
Catalog number/FG Item Number: 263J-SS-SET 1. Specific Lots with Bowl Cracks (110 Distributed) Lot number: 0718027-0119029-08170710119003; 0718027-0818086-08170711018045; 0718027-1118031-08170711118019; 0718027-1118081-08170710119084 2. Missing instructions to check for cracks in IFU All products/lost manufactured between October 2014 and October 2019.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Japan DISP SET, 263J, 205J, HAR - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma recalled?

Yes. FDA published recall Z-0906-2020 on February 5, 2020. Reported hazard: Potential to develop leaks at the inner core of Cell Saver® 5/5+ and Cell Saver® Elite®/Elite® +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.

What should I do if I have this product?

Voluntary: Firm initiated · Letter