Firebird Spinal Fixation System, Parallel Rod Connector (52-6800) and Axial Rod Connector (52-6700). Product Usage: The connector assemblies allow for rod components to be connected side-to-side, or end-to-end and eliminate the need of rebuilding the entire system with a longer rod in order to include the new vertebral levels.
Published March 6, 2013 · Recall initiated: February 1, 2013
Summary
FDA records medical device Class II recall Z-0896-2013, published March 6, 2013 and initiated February 1, 2013. Recalling firm: Orthofix, Inc; Status: Terminated; Reported quantity: 565 total (287 PN 52-6800 and 278 PN 52- 6700); Lot: Catalog Numbers: 52-6800 & 52-6700; All Lot Numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is a possibility that the Set Screw (p/n 90-2923) sub-component of the Parallel Rod Connector Assembly (p/n 52-6800) & Axial Rod Connector (p/n 52-6700) may crack during final tightening. Orthofix Inc received six (6) customer complaints related to set screw cracking between November 15, 2010 and December 2012.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Orthofix, Inc
- Distributor
- Worldwidewide Distribution - USA Nationwide including Puerto Rico and the countries of: Australia, Lebanon, Germany, Spain, Sweden and Israel.
- Lot
- Catalog Numbers: 52-6800 & 52-6700; All Lot Numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDARival Reduce Kit Locking and Non-Locking Screw D 2.0MM, D 2.7MM, Sterile, Rx Only, ORTHOFIX S.r.r. Italy onJune 13, 2018
- FDAY-Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) Small (003-C-01001); b)June 13, 2018
- FDAL-Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Left or Right (Part Number): a) LeftJune 13, 2018
- FDAFirebird NXG Spinal Fixation System Decorticating Planer, Catalog #36-1334, non-sterile, Rx. .A DecorticatingSeptember 20, 2017
- FDAOrthofix ISKD Limb Lengthener. ISKD - Intramedullary Skeletal Kinetic Distractor. The ISKD is intended forMarch 6, 2013
- FDAIngenuity TF PET/CT. Diagnostic imaging systemMarch 6, 2013
Questions about this recall
Was Firebird Spinal Fixation System, Parallel Rod Connector (52-6800) and Axial Rod Connector (52-6700). Product Usage: The connector assemblies allow for rod components to be connected side-to-side, or end-to-end and eliminate the need of rebuilding the entire system with a longer rod in order to include the new vertebral levels. recalled?
Yes. FDA published recall Z-0896-2013 on March 6, 2013. Reported hazard: There is a possibility that the Set Screw (p/n 90-2923) sub-component of the Parallel Rod Connector Assembly (p/n 52-6800) & Axial Rod Connector (p/n 52-6700) may crack during final tightening. Orthofix Inc received six (6) customer complaints related to set screw cracking between November 15, 2010 and December 2012.
What should I do if I have this product?
Voluntary: Firm initiated · Letter