Firebird Spinal Fixation System, Parallel Rod Connector (52-6800) and Axial Rod Connector (52-6700). Product Usage: The connector assemblies allow for rod components to be connected side-to-side, or end-to-end and eliminate the need of rebuilding the entire system with a longer rod in order to include the new vertebral levels.

Published March 6, 2013 · Recall initiated: February 1, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There is a possibility that the Set Screw (p/n 90-2923) sub-component of the Parallel Rod Connector Assembly (p/n 52-6800) & Axial Rod Connector (p/n 52-6700) may crack during final tightening. Orthofix Inc received six (6) customer complaints related to set screw cracking between November 15, 2010 and December 2012.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Orthofix, Inc
Distributor
Worldwidewide Distribution - USA Nationwide including Puerto Rico and the countries of: Australia, Lebanon, Germany, Spain, Sweden and Israel.
Lot
Catalog Numbers: 52-6800 & 52-6700; All Lot Numbers
Risk classification
Class II
Area (FDA)
device
Country
United States