GE Healthcare Care Plus Hood Field Replacement Unit (FRU) kit, Part Numbers: 6600-0047-851, 6600-0033-850, 6600-0081-850, 6600-0961-500, 6600-0083-850
Published February 21, 2024 · Recall initiated: December 26, 2023
Summary
FDA records medical device Class I recall Z-0959-2024, published February 21, 2024 and initiated December 26, 2023. Recalling firm: Wipro GE Healthcare Private Ltd.; Status: Ongoing; Reported quantity: 1465 units; Lot: ALL SERIAL NUMBERS. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus® Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Wipro GE Healthcare Private Ltd.
- Distributor
- Worldwide distribution.
- Lot
- ALL SERIAL NUMBERS
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- India
Related recalls
- FDAGE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5, Part NumberOctober 29, 2025
- FDAGE Panda iRES Warmer, Model/Catalog Numbers: 1) M1112198-01045518; 2) M1112198-01045579; 3) M1112198-01047148;October 2, 2024
- FDAGE Giraffe Warmer, Model/Catalog Number M1118179-01090979. Infant radiant warmers provide infrared heat in aOctober 2, 2024
- FDAGE Healthcare Lullaby Incubator Model Hood FRU kit, Part Numbers: M1169566, M1160659, M1206259, M1206260,February 21, 2024
- FDAGE Healthcare neonatal incubator, models: a) Lullaby Incubator, b) Lullaby Incubator XP, c) Lullaby IncubatorFebruary 21, 2024
- FDAGE Healthcare Care Plus Hood Door FRU kit, Part Numbers: 6600-0082-850, 6600-0504-800, 6600-0148-800, andFebruary 21, 2024
Questions about this recall
Was GE Healthcare Care Plus Hood Field Replacement Unit (FRU) kit, Part Numbers: 6600-0047-851, 6600-0033-850, 6600-0081-850, 6600-0961-500, 6600-0083-850 recalled?
Yes. FDA published recall Z-0959-2024 on February 21, 2024. Reported hazard: GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus® Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.
What should I do if I have this product?
Voluntary: Firm initiated · Letter