GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5, Part Number 5867474-03; Medical Display, cathode ray tube

Published October 29, 2025 · Recall initiated: September 5, 2025

Summary

FDA records medical device Class II recall Z-0214-2026, published October 29, 2025 and initiated September 5, 2025. Recalling firm: Wipro GE Healthcare Private Ltd.; Status: Ongoing; Reported quantity: 2365 units; Lot: GTIN 00195278343383, Serial Numbers: STV25225456SA, STV25225455SA, STV25266464SA, STV25235817SA, STV25235816SA, STV25286682SA, STV25225594SA, STV25235792SA, STV25235774SA, STV25225550SA, STV25235626SA, STV25235625SA, STV25184592SA, STV25235628SA, STV25235631SA, STV25184595SA, STV25235629SA, STV25235632SA, STV25235630SA, STV25276515SA, STV25225510SA, STV25276516SA, STV25225505SA, STV25225508SA, STV25225509SA, STV25225507SA, STV25225506SA, STV25113369SA, STV25204986SA, STV25205160SA, STV25205151SA, STV25205161SA, STV25062736SA, STV25256133SA, STV25225512SA, STV25296702SA, STV25296701SA, STV25296748SA, STV25256214SA, STV25225511SA, STV25256215SA, STV25256199SA, STV25256216SA, STV25256202SA, STV25225497SA, STV25225514SA, STV25225498SA, STV25256131SA, STV25256129SA, STV25256128SA, STV25296747SA, STV25256132SA, STV25256362SA, STV25225494SA, STV25225513SA, STV25296697SA, STV25256153SA, STV25296745SA, STV25256130SA, STV25256134SA, STV25256198SA, STV25296749SA, STV25296698SA, STV25256112SA, STV25256111SA, STV25225495SA, STV25256127SA, STV25256197SA, STV25296700SA, STV25296699SA, STV23440159SA, STV23440160SA, STV23440161SA, STV23440162SA, STV23440163SA, STV23440164SA, STV23440165SA, STV23440166SA, STV23440167SA, STV23440169SA, STV23440170SA, STV23440172SA, STV23440173SA, STV23440174SA, STV23440175SA, STV23440176SA, STV23440178SA, STV23440179SA, STV23440180SA, STV23440185SA, STV23440186SA, STV23440187SA, STV23440188SA, STV24250149SA, STV24250150SA, STV24250151SA, STV23440171SA, STV25184584SA, STV25174487SA, STV25256226SA, STV25256227SA, STV25256229SA, STV25256231SA, STV25256232SA, STV25256233SA, STV25256242SA, STV25256243SA, STV25256245SA, STV25256246SA, STV25256247SA, STV25256250SA, STV25256257SA, STV25256258SA, STV23450102SA, STV23450100SA, STV23450101SA, STV23450104SA, STV23450110SA, STV23450108SA, STV23450099SA, STV23450109SA, STV25266498SA, STV25266499SA, STV25184700SA, STV25194774SA, STV25194847SA, STV25194865SA, STV25194866SA, STV25194864SA, STV25194874SA, STV25194852SA, STV25194853SA, STV25194861SA, STV25194863SA, STV25194873SA, STV25194857SA, STV25194867SA, STV25225395SA, STV25225384SA, STV25225377SA, STV25266497SA, STV25225385SA, STV25225383SA, STV25225357SA, STV25225389SA, STV25225396SA, STV25225378SA, STV25225388SA, STV25225393SA, STV25225381SA, STV25225604SA, STV25296707SA, STV25276543SA, STV25276540SA, STV25276545SA, STV25276505SA, STV25276544SA, STV25276524SA, STV25276506SA, STV25276542SA, STV25276541SA, STV25276539SA, STV25276520SA, STV25194979SA, STV25184651SA, STV25215224SA, STV25215244SA, STV25215324SA, STV25215218SA, STV25215240SA, STV25215241SA, STV25215217SA, STV25215220SA, STV25215323SA, STV25215238SA, STV25215307SA, STV25215222SA, STV25215237SA, STV25215243SA, STV25215221SA, STV25215304SA, STV25215306SA, STV25215247SA, STV25215223SA, STV25215248SA, STV25215250SA, STV25256224SA, STV25215239SA, STV25215249SA, STV25215245SA, STV25215305SA, STV25215246SA, STV24250041SA, STV25245906SA, STV25245947SA, STV25245944SA, STV25245926SA, STV25245924SA, STV25245907SA, STV25225418SA, STV25245942SA, STV25245949SA, STV25245894SA, STV25245945SA, STV25276547SA, STV25276548SA, STV25276554SA, STV25276555SA, STV23280024SA, STV23280025SA, STV25052476SA, STV25052477SA, STV25052471SA, STV25225618SA, STV25225609SA, STV25225610SA, STV25225587SA, STV25225588SA, STV25296727SA, STV25215274SA, STV25215269SA, STV25215259SA, STV25215256SA, STV25215235SA, STV25266496SA, STV25215232SA, STV25215234SA, STV25215229SA, STV25215227SA, STV25215228SA, STV25215266SA, STV25215260SA, STV25215272SA, STV25215273SA, STV25215257SA, STV25215226SA, STV25215236SA, STV25215275SA, STV25215277SA, STV25215270SA, STV25215225SA, STV25215267SA, STV25215231SA, STV25215268SA, STV25215258SA, STV25215233SA, STV25215271SA, STV25215276SA, STV25215251SA, STV25215230SA, STV25205213SA, STV25205205SA, STV25205209SA, STV25205208SA, STV25205212SA, STV25205211SA, STV25205210SA, STV25205207SA, STV25205215SA, STV25205214SA, STV25205206SA, STV25296713SA, STV25286676SA, STV25286672SA, STV. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Wipro GE Healthcare Private Ltd.
Distributor
Worldwide distribution.
Lot
GTIN 00195278343383, Serial Numbers: STV25225456SA, STV25225455SA, STV25266464SA, STV25235817SA, STV25235816SA, STV25286682SA, STV25225594SA, STV25235792SA, STV25235774SA, STV25225550SA, STV25235626SA, STV25235625SA, STV25184592SA, STV25235628SA, STV25235631SA, STV25184595SA, STV25235629SA, STV25235632SA, STV25235630SA, STV25276515SA, STV25225510SA, STV25276516SA, STV25225505SA, STV25225508SA, STV25225509SA, STV25225507SA, STV25225506SA, STV25113369SA, STV25204986SA, STV25205160SA, STV25205151SA, STV25205161SA, STV25062736SA, STV25256133SA, STV25225512SA, STV25296702SA, STV25296701SA, STV25296748SA, STV25256214SA, STV25225511SA, STV25256215SA, STV25256199SA, STV25256216SA, STV25256202SA, STV25225497SA, STV25225514SA, STV25225498SA, STV25256131SA, STV25256129SA, STV25256128SA, STV25296747SA, STV25256132SA, STV25256362SA, STV25225494SA, STV25225513SA, STV25296697SA, STV25256153SA, STV25296745SA, STV25256130SA, STV25256134SA, STV25256198SA, STV25296749SA, STV25296698SA, STV25256112SA, STV25256111SA, STV25225495SA, STV25256127SA, STV25256197SA, STV25296700SA, STV25296699SA, STV23440159SA, STV23440160SA, STV23440161SA, STV23440162SA, STV23440163SA, STV23440164SA, STV23440165SA, STV23440166SA, STV23440167SA, STV23440169SA, STV23440170SA, STV23440172SA, STV23440173SA, STV23440174SA, STV23440175SA, STV23440176SA, STV23440178SA, STV23440179SA, STV23440180SA, STV23440185SA, STV23440186SA, STV23440187SA, STV23440188SA, STV24250149SA, STV24250150SA, STV24250151SA, STV23440171SA, STV25184584SA, STV25174487SA, STV25256226SA, STV25256227SA, STV25256229SA, STV25256231SA, STV25256232SA, STV25256233SA, STV25256242SA, STV25256243SA, STV25256245SA, STV25256246SA, STV25256247SA, STV25256250SA, STV25256257SA, STV25256258SA, STV23450102SA, STV23450100SA, STV23450101SA, STV23450104SA, STV23450110SA, STV23450108SA, STV23450099SA, STV23450109SA, STV25266498SA, STV25266499SA, STV25184700SA, STV25194774SA, STV25194847SA, STV25194865SA, STV25194866SA, STV25194864SA, STV25194874SA, STV25194852SA, STV25194853SA, STV25194861SA, STV25194863SA, STV25194873SA, STV25194857SA, STV25194867SA, STV25225395SA, STV25225384SA, STV25225377SA, STV25266497SA, STV25225385SA, STV25225383SA, STV25225357SA, STV25225389SA, STV25225396SA, STV25225378SA, STV25225388SA, STV25225393SA, STV25225381SA, STV25225604SA, STV25296707SA, STV25276543SA, STV25276540SA, STV25276545SA, STV25276505SA, STV25276544SA, STV25276524SA, STV25276506SA, STV25276542SA, STV25276541SA, STV25276539SA, STV25276520SA, STV25194979SA, STV25184651SA, STV25215224SA, STV25215244SA, STV25215324SA, STV25215218SA, STV25215240SA, STV25215241SA, STV25215217SA, STV25215220SA, STV25215323SA, STV25215238SA, STV25215307SA, STV25215222SA, STV25215237SA, STV25215243SA, STV25215221SA, STV25215304SA, STV25215306SA, STV25215247SA, STV25215223SA, STV25215248SA, STV25215250SA, STV25256224SA, STV25215239SA, STV25215249SA, STV25215245SA, STV25215305SA, STV25215246SA, STV24250041SA, STV25245906SA, STV25245947SA, STV25245944SA, STV25245926SA, STV25245924SA, STV25245907SA, STV25225418SA, STV25245942SA, STV25245949SA, STV25245894SA, STV25245945SA, STV25276547SA, STV25276548SA, STV25276554SA, STV25276555SA, STV23280024SA, STV23280025SA, STV25052476SA, STV25052477SA, STV25052471SA, STV25225618SA, STV25225609SA, STV25225610SA, STV25225587SA, STV25225588SA, STV25296727SA, STV25215274SA, STV25215269SA, STV25215259SA, STV25215256SA, STV25215235SA, STV25266496SA, STV25215232SA, STV25215234SA, STV25215229SA, STV25215227SA, STV25215228SA, STV25215266SA, STV25215260SA, STV25215272SA, STV25215273SA, STV25215257SA, STV25215226SA, STV25215236SA, STV25215275SA, STV25215277SA, STV25215270SA, STV25215225SA, STV25215267SA, STV25215231SA, STV25215268SA, STV25215258SA, STV25215233SA, STV25215271SA, STV25215276SA, STV25215251SA, STV25215230SA, STV25205213SA, STV25205205SA, STV25205209SA, STV25205208SA, STV25205212SA, STV25205211SA, STV25205210SA, STV25205207SA, STV25205215SA, STV25205214SA, STV25205206SA, STV25296713SA, STV25286676SA, STV25286672SA, STV
Risk classification
Class II
Area (FDA)
device
Country
India

Related recalls

Questions about this recall

Was GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5, Part Number 5867474-03; Medical Display, cathode ray tube recalled?

Yes. FDA published recall Z-0214-2026 on October 29, 2025. Reported hazard: Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.

What should I do if I have this product?

Voluntary: Firm initiated · Letter