GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5, Part Number 5867474-03; Medical Display, cathode ray tube
Published October 29, 2025 · Recall initiated: September 5, 2025
Summary
FDA records medical device Class II recall Z-0214-2026, published October 29, 2025 and initiated September 5, 2025. Recalling firm: Wipro GE Healthcare Private Ltd.; Status: Ongoing; Reported quantity: 2365 units; Lot: GTIN 00195278343383, Serial Numbers: STV25225456SA, STV25225455SA, STV25266464SA, STV25235817SA, STV25235816SA, STV25286682SA, STV25225594SA, STV25235792SA, STV25235774SA, STV25225550SA, STV25235626SA, STV25235625SA, STV25184592SA, STV25235628SA, STV25235631SA, STV25184595SA, STV25235629SA, STV25235632SA, STV25235630SA, STV25276515SA, STV25225510SA, STV25276516SA, STV25225505SA, STV25225508SA, STV25225509SA, STV25225507SA, STV25225506SA, STV25113369SA, STV25204986SA, STV25205160SA, STV25205151SA, STV25205161SA, STV25062736SA, STV25256133SA, STV25225512SA, STV25296702SA, STV25296701SA, STV25296748SA, STV25256214SA, STV25225511SA, STV25256215SA, STV25256199SA, STV25256216SA, STV25256202SA, STV25225497SA, STV25225514SA, STV25225498SA, STV25256131SA, STV25256129SA, STV25256128SA, STV25296747SA, STV25256132SA, STV25256362SA, STV25225494SA, STV25225513SA, STV25296697SA, STV25256153SA, STV25296745SA, STV25256130SA, STV25256134SA, STV25256198SA, STV25296749SA, STV25296698SA, STV25256112SA, STV25256111SA, STV25225495SA, STV25256127SA, STV25256197SA, STV25296700SA, STV25296699SA, STV23440159SA, STV23440160SA, STV23440161SA, STV23440162SA, STV23440163SA, STV23440164SA, STV23440165SA, STV23440166SA, STV23440167SA, STV23440169SA, STV23440170SA, STV23440172SA, STV23440173SA, STV23440174SA, STV23440175SA, STV23440176SA, STV23440178SA, STV23440179SA, STV23440180SA, STV23440185SA, STV23440186SA, STV23440187SA, STV23440188SA, STV24250149SA, STV24250150SA, STV24250151SA, STV23440171SA, STV25184584SA, STV25174487SA, STV25256226SA, STV25256227SA, STV25256229SA, STV25256231SA, STV25256232SA, STV25256233SA, STV25256242SA, STV25256243SA, STV25256245SA, STV25256246SA, STV25256247SA, STV25256250SA, STV25256257SA, STV25256258SA, STV23450102SA, STV23450100SA, STV23450101SA, STV23450104SA, STV23450110SA, STV23450108SA, STV23450099SA, STV23450109SA, STV25266498SA, STV25266499SA, STV25184700SA, STV25194774SA, STV25194847SA, STV25194865SA, STV25194866SA, STV25194864SA, STV25194874SA, STV25194852SA, STV25194853SA, STV25194861SA, STV25194863SA, STV25194873SA, STV25194857SA, STV25194867SA, STV25225395SA, STV25225384SA, STV25225377SA, STV25266497SA, STV25225385SA, STV25225383SA, STV25225357SA, STV25225389SA, STV25225396SA, STV25225378SA, STV25225388SA, STV25225393SA, STV25225381SA, STV25225604SA, STV25296707SA, STV25276543SA, STV25276540SA, STV25276545SA, STV25276505SA, STV25276544SA, STV25276524SA, STV25276506SA, STV25276542SA, STV25276541SA, STV25276539SA, STV25276520SA, STV25194979SA, STV25184651SA, STV25215224SA, STV25215244SA, STV25215324SA, STV25215218SA, STV25215240SA, STV25215241SA, STV25215217SA, STV25215220SA, STV25215323SA, STV25215238SA, STV25215307SA, STV25215222SA, STV25215237SA, STV25215243SA, STV25215221SA, STV25215304SA, STV25215306SA, STV25215247SA, STV25215223SA, STV25215248SA, STV25215250SA, STV25256224SA, STV25215239SA, STV25215249SA, STV25215245SA, STV25215305SA, STV25215246SA, STV24250041SA, STV25245906SA, STV25245947SA, STV25245944SA, STV25245926SA, STV25245924SA, STV25245907SA, STV25225418SA, STV25245942SA, STV25245949SA, STV25245894SA, STV25245945SA, STV25276547SA, STV25276548SA, STV25276554SA, STV25276555SA, STV23280024SA, STV23280025SA, STV25052476SA, STV25052477SA, STV25052471SA, STV25225618SA, STV25225609SA, STV25225610SA, STV25225587SA, STV25225588SA, STV25296727SA, STV25215274SA, STV25215269SA, STV25215259SA, STV25215256SA, STV25215235SA, STV25266496SA, STV25215232SA, STV25215234SA, STV25215229SA, STV25215227SA, STV25215228SA, STV25215266SA, STV25215260SA, STV25215272SA, STV25215273SA, STV25215257SA, STV25215226SA, STV25215236SA, STV25215275SA, STV25215277SA, STV25215270SA, STV25215225SA, STV25215267SA, STV25215231SA, STV25215268SA, STV25215258SA, STV25215233SA, STV25215271SA, STV25215276SA, STV25215251SA, STV25215230SA, STV25205213SA, STV25205205SA, STV25205209SA, STV25205208SA, STV25205212SA, STV25205211SA, STV25205210SA, STV25205207SA, STV25205215SA, STV25205214SA, STV25205206SA, STV25296713SA, STV25286676SA, STV25286672SA, STV. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Wipro GE Healthcare Private Ltd.
- Distributor
- Worldwide distribution.
- Lot
- GTIN 00195278343383, Serial Numbers: STV25225456SA, STV25225455SA, STV25266464SA, STV25235817SA, STV25235816SA, STV25286682SA, STV25225594SA, STV25235792SA, STV25235774SA, STV25225550SA, STV25235626SA, STV25235625SA, STV25184592SA, STV25235628SA, STV25235631SA, STV25184595SA, STV25235629SA, STV25235632SA, STV25235630SA, STV25276515SA, STV25225510SA, STV25276516SA, STV25225505SA, STV25225508SA, STV25225509SA, STV25225507SA, STV25225506SA, STV25113369SA, STV25204986SA, STV25205160SA, STV25205151SA, STV25205161SA, STV25062736SA, STV25256133SA, STV25225512SA, STV25296702SA, STV25296701SA, STV25296748SA, STV25256214SA, STV25225511SA, STV25256215SA, STV25256199SA, STV25256216SA, STV25256202SA, STV25225497SA, STV25225514SA, STV25225498SA, STV25256131SA, STV25256129SA, STV25256128SA, STV25296747SA, STV25256132SA, STV25256362SA, STV25225494SA, STV25225513SA, STV25296697SA, STV25256153SA, STV25296745SA, STV25256130SA, STV25256134SA, STV25256198SA, STV25296749SA, STV25296698SA, STV25256112SA, STV25256111SA, STV25225495SA, STV25256127SA, STV25256197SA, STV25296700SA, STV25296699SA, STV23440159SA, STV23440160SA, STV23440161SA, STV23440162SA, STV23440163SA, STV23440164SA, STV23440165SA, STV23440166SA, STV23440167SA, STV23440169SA, STV23440170SA, STV23440172SA, STV23440173SA, STV23440174SA, STV23440175SA, STV23440176SA, STV23440178SA, STV23440179SA, STV23440180SA, STV23440185SA, STV23440186SA, STV23440187SA, STV23440188SA, STV24250149SA, STV24250150SA, STV24250151SA, STV23440171SA, STV25184584SA, STV25174487SA, STV25256226SA, STV25256227SA, STV25256229SA, STV25256231SA, STV25256232SA, STV25256233SA, STV25256242SA, STV25256243SA, STV25256245SA, STV25256246SA, STV25256247SA, STV25256250SA, STV25256257SA, STV25256258SA, STV23450102SA, STV23450100SA, STV23450101SA, STV23450104SA, STV23450110SA, STV23450108SA, STV23450099SA, STV23450109SA, STV25266498SA, STV25266499SA, STV25184700SA, STV25194774SA, STV25194847SA, STV25194865SA, STV25194866SA, STV25194864SA, STV25194874SA, STV25194852SA, STV25194853SA, STV25194861SA, STV25194863SA, STV25194873SA, STV25194857SA, STV25194867SA, STV25225395SA, STV25225384SA, STV25225377SA, STV25266497SA, STV25225385SA, STV25225383SA, STV25225357SA, STV25225389SA, STV25225396SA, STV25225378SA, STV25225388SA, STV25225393SA, STV25225381SA, STV25225604SA, STV25296707SA, STV25276543SA, STV25276540SA, STV25276545SA, STV25276505SA, STV25276544SA, STV25276524SA, STV25276506SA, STV25276542SA, STV25276541SA, STV25276539SA, STV25276520SA, STV25194979SA, STV25184651SA, STV25215224SA, STV25215244SA, STV25215324SA, STV25215218SA, STV25215240SA, STV25215241SA, STV25215217SA, STV25215220SA, STV25215323SA, STV25215238SA, STV25215307SA, STV25215222SA, STV25215237SA, STV25215243SA, STV25215221SA, STV25215304SA, STV25215306SA, STV25215247SA, STV25215223SA, STV25215248SA, STV25215250SA, STV25256224SA, STV25215239SA, STV25215249SA, STV25215245SA, STV25215305SA, STV25215246SA, STV24250041SA, STV25245906SA, STV25245947SA, STV25245944SA, STV25245926SA, STV25245924SA, STV25245907SA, STV25225418SA, STV25245942SA, STV25245949SA, STV25245894SA, STV25245945SA, STV25276547SA, STV25276548SA, STV25276554SA, STV25276555SA, STV23280024SA, STV23280025SA, STV25052476SA, STV25052477SA, STV25052471SA, STV25225618SA, STV25225609SA, STV25225610SA, STV25225587SA, STV25225588SA, STV25296727SA, STV25215274SA, STV25215269SA, STV25215259SA, STV25215256SA, STV25215235SA, STV25266496SA, STV25215232SA, STV25215234SA, STV25215229SA, STV25215227SA, STV25215228SA, STV25215266SA, STV25215260SA, STV25215272SA, STV25215273SA, STV25215257SA, STV25215226SA, STV25215236SA, STV25215275SA, STV25215277SA, STV25215270SA, STV25215225SA, STV25215267SA, STV25215231SA, STV25215268SA, STV25215258SA, STV25215233SA, STV25215271SA, STV25215276SA, STV25215251SA, STV25215230SA, STV25205213SA, STV25205205SA, STV25205209SA, STV25205208SA, STV25205212SA, STV25205211SA, STV25205210SA, STV25205207SA, STV25205215SA, STV25205214SA, STV25205206SA, STV25296713SA, STV25286676SA, STV25286672SA, STV
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- India
Related recalls
- FDASalem Sump PVC Tubes: Product Code/Product Name: 1180264408 Salem Sump" PVC Gastroduodenal Tube, 10 Fr/Ch (3.3October 29, 2025
- FDAShanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 550 MD, Rx OnlyOctober 29, 2025
- FDAGE Giraffe Warmer, Model/Catalog Number M1118179-01090979. Infant radiant warmers provide infrared heat in aOctober 2, 2024
- FDAGE Panda iRES Warmer, Model/Catalog Numbers: 1) M1112198-01045518; 2) M1112198-01045579; 3) M1112198-01047148;October 2, 2024
- FDAGE Healthcare Lullaby Incubator Model Hood Door FRU kit, Part Numbers: M1169767, M1160674, M1187512, M1206646,February 21, 2024
- FDAGE Healthcare Care Plus Hood Field Replacement Unit (FRU) kit, Part Numbers: 6600-0047-851, 6600-0033-850,February 21, 2024
Questions about this recall
Was GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5, Part Number 5867474-03; Medical Display, cathode ray tube recalled?
Yes. FDA published recall Z-0214-2026 on October 29, 2025. Reported hazard: Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.
What should I do if I have this product?
Voluntary: Firm initiated · Letter