GE Giraffe Warmer, Model/Catalog Number M1118179-01090979. Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment.
Published October 2, 2024 · Recall initiated: August 9, 2024
Summary
FDA records medical device Class II recall Z-3080-2024, published October 2, 2024 and initiated August 9, 2024. Recalling firm: Wipro GE Healthcare Private Ltd.; Status: Ongoing; Reported quantity: 1; Lot: Serial Number GBW24230381SA. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
GE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the product design. However, because the electrical safety testing in manufacturing was incomplete for these devices, there is the potential for leakage current to exceed IEC 60601 limits.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Wipro GE Healthcare Private Ltd.
- Distributor
- US and Australia, Denmark, France, Germany, Ireland, ITALY, New Zealand, Poland, Portugal, Slovenia, Spain, SWITZERLAND, United Arab Emirates, United Kingdom.
- Lot
- Serial Number GBW24230381SA
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- India
Related recalls
- FDAGE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5, Part NumberOctober 29, 2025
- FDAGE Panda iRES Warmer, Model/Catalog Numbers: 1) M1112198-01045518; 2) M1112198-01045579; 3) M1112198-01047148;October 2, 2024
- FDABacalaitos Criollos, CODFISH FRYING MIX, 7 oz., packaged in a plastic bag, 24 packages (units) per cases of 7October 2, 2024
- FDAMiso Katsuo Ninniku (Pickled Garlic), Net WT 5.2 OZ (147g) Products of China; Imported By Niitakaya U.S.A.October 2, 2024
- FDAGE Healthcare Lullaby Incubator Model Hood Door FRU kit, Part Numbers: M1169767, M1160674, M1187512, M1206646,February 21, 2024
- FDAGE Healthcare Care Plus Hood Field Replacement Unit (FRU) kit, Part Numbers: 6600-0047-851, 6600-0033-850,February 21, 2024
Questions about this recall
Was GE Giraffe Warmer, Model/Catalog Number M1118179-01090979. Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment. recalled?
Yes. FDA published recall Z-3080-2024 on October 2, 2024. Reported hazard: GE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the product design. However, because the electrical safety testing in manufacturing was incomplete for these devices, there is the potential for leakage current to exceed IEC 60601 limits.
What should I do if I have this product?
Voluntary: Firm initiated · Letter