GE Healthcare Lullaby Incubator Model Hood Door FRU kit, Part Numbers: M1169767, M1160674, M1187512, M1206646, M1166822
Published February 21, 2024 · Recall initiated: December 26, 2023
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus¿ Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Wipro GE Healthcare Private Ltd.
- Distributor
- Worldwide distribution.
- Lot
- ALL SERIAL NUMBERS
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- India