Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509340 Software Version: N/A Product Description: Affixus 9 mm, Length 340 mm, Right, Trochanteric Nail Component: N/A
Published December 31, 2025 · Recall initiated: December 2, 2025
Summary
FDA records medical device Class II recall Z-0968-2026, published December 31, 2025 and initiated December 2, 2025. Recalling firm: Zimmer, Inc.; Status: Ongoing; Reported quantity: 22; Lot: Lot Code: Model No 815509340 ; UDI-DI (01)00887868143533(17)340822(10)66717552 ; Lot Number 66717552 Model No 815509340 ; UDI-DI (01)00887868143533(17)340826(10)66884562 ; Lot Number 66884562. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer, Inc.
- Distributor
- US Nationwide distribution in the states of AL, AZ, CA, FL, IN, KY, MA, MO, MT, NC, NM, RI, TX, UT.
- Lot
- Lot Code: Model No 815509340 ; UDI-DI (01)00887868143533(17)340822(10)66717552 ; Lot Number 66717552 Model No 815509340 ; UDI-DI (01)00887868143533(17)340826(10)66884562 ; Lot Number 66884562
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOff-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 110040240July 1, 2026
- FDABrand Name: Persona Revision Product Name: Persona Revision® Trabecular Metal® Femoral Distal AugmentJuly 1, 2026
- FDAOff-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444July 1, 2026
- FDAOff-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number:July 1, 2026
- FDALifeShield Drug Library Management (DLM); LifeShield Infusion Safety Software Suite v2.2 List Number 17003-02December 31, 2025
- FDABrand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral NailsDecember 31, 2025
Questions about this recall
Was Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509340 Software Version: N/A Product Description: Affixus 9 mm, Length 340 mm, Right, Trochanteric Nail Component: N/A recalled?
Yes. FDA published recall Z-0968-2026 on December 31, 2025. Reported hazard: Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
What should I do if I have this product?
Voluntary: Firm initiated · Letter