Specialist 2 IM Rod, Non-Sterile and associated Surgical Techniques: Sigma HP Instruments Classic Surgical Technique,0612-89-510 Sigma HP Instruments Balanced Surgical Technique, 0612-88-510 Sigma HP Instruments Fixed Reference Surgical Technique, 0612-87-510 PFC Sigma Knee Systems Primary Cruciate-Retaining & Cruciate-Substituting Procedure Surgical Technique,SP2-007 Product Usage: The SP2 IM Rod is used in both primary and revision Sigma knee procedures to align the femoral locating device and distal femoral cutting block. It is also used with the IM tibial resection. This rod is included in both the SP2 and HP instrument kits.
Published March 27, 2013 · Recall initiated: February 5, 2013
Summary
FDA records medical device Class II recall Z-0971-2013, published March 27, 2013 and initiated February 5, 2013. Recalling firm: DePuy Orthopaedics, Inc.; Status: Terminated; Reported quantity: 4000; 2600 in USA; Lot: Product Code 96-6120 Label code / Etch code C3JHN4 H0808 C3JHS4 H0908 C3JHF4 H1008 C4GA54 C4GCC4 C4GBT4 C52F74 H1108 C52GV4 H1208 C52GL4 C67N14 C98CS4 H0109 C98BF4 DE5P34 H0309 DE5RP4 DF4H44 DG9LK4 H0409 DG9L64 DJ5E34 DK3E34 H0509 DK3FE4 EB5FV4 H0210 D95AN4 EB5GH4 H0310 EC9JY4 EF4DJ4 EJ7AP4 H0410 ES2G64 H0510 EJ7A34 ES2HA4 H0610 ES2HY4 H0710 EX5L44 EX5MS4 H0810 E2SD44 H0910 FA4G94 H0211 FD8MP4 H0311 FH8JA4 FH8JX4 H0611 TBACC TBACC TBACZ TBACZ FJ4E74 TBAGG. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential for the IM rod to break, leaving fragments in the patient. Surgical Techniques guidance has been updated until design changes can be made to reduce likelihood of potential failure.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- DePuy Orthopaedics, Inc.
- Distributor
- Worldwide Distribution - USA Nationwide.
- Lot
- Product Code 96-6120 Label code / Etch code C3JHN4 H0808 C3JHS4 H0908 C3JHF4 H1008 C4GA54 C4GCC4 C4GBT4 C52F74 H1108 C52GV4 H1208 C52GL4 C67N14 C98CS4 H0109 C98BF4 DE5P34 H0309 DE5RP4 DF4H44 DG9LK4 H0409 DG9L64 DJ5E34 DK3E34 H0509 DK3FE4 EB5FV4 H0210 D95AN4 EB5GH4 H0310 EC9JY4 EF4DJ4 EJ7AP4 H0410 ES2G64 H0510 EJ7A34 ES2HA4 H0610 ES2HY4 H0710 EX5L44 EX5MS4 H0810 E2SD44 H0910 FA4G94 H0211 FD8MP4 H0311 FH8JA4 FH8JX4 H0611 TBACC TBACC TBACZ TBACZ FJ4E74 TBAGG
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAATTUNE AFFIXIUM Cementless Fixed Bearing Knee with 3DP Technology, Size 6. Component of the ATTUNE Total KneeApril 10, 2024
- FDAAttune Revision Limb Preservation System (LPS) Tibial Insert XSM 12MM, Part Number 151760212March 29, 2023
- FDAAttune Revision Limb Preservation System (LPS) Tibial Insert SM 12MM, Part Number 151760312March 29, 2023
- FDAATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the bone cuts in preparationJuly 27, 2022
- FDACR MD1.0 General Set, x-ray imager image plate. CR 10-X Image Plate 35CM X 43CM. An image plate, contained inMarch 27, 2013
- FDAReClaim Reamer Extension; Product code 2975-00-500, used with Reclaim Modular Revision Hip Prosthesis System.March 27, 2013
Questions about this recall
Was Specialist 2 IM Rod, Non-Sterile and associated Surgical Techniques: Sigma HP Instruments Classic Surgical Technique,0612-89-510 Sigma HP Instruments Balanced Surgical Technique, 0612-88-510 Sigma HP Instruments Fixed Reference Surgical Technique, 0612-87-510 PFC Sigma Knee Systems Primary Cruciate-Retaining & Cruciate-Substituting Procedure Surgical Technique,SP2-007 Product Usage: The SP2 IM Rod is used in both primary and revision Sigma knee procedures to align the femoral locating device and distal femoral cutting block. It is also used with the IM tibial resection. This rod is included in both the SP2 and HP instrument kits. recalled?
Yes. FDA published recall Z-0971-2013 on March 27, 2013. Reported hazard: Potential for the IM rod to break, leaving fragments in the patient. Surgical Techniques guidance has been updated until design changes can be made to reduce likelihood of potential failure.
What should I do if I have this product?
Voluntary: Firm initiated · Letter