Back to search
FDAFDA · Food and drugs·Recall number Z-1015-2017

Infant Warmer System (IWS)

Published January 18, 2017 · Recall initiated: August 4, 2016

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

GE Healthcare has recently become aware of a potential safety issue related to loose screws in the Heater Head of the Infant Warmer System (IWS). Hot screws from the Heater Head of the IWS could fall onto the bed if the Heater Head assembly has been improperly serviced. This situation can be clinically hazardous because thermal injury to a patient could result. Two injuries have been reported as a result of this issue.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
GE Healthcare, LLC
Distributor
This product has been distributed domestically and internationally .
Lot
Medical device listing number: D103053, D103054, D103056, D103057, D183529
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card