regard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator. Convenience kit to be used in Labor and Delivery.
Published February 12, 2025 · Recall initiated: December 26, 2024
Summary
FDA records medical device Class I recall Z-1047-2025, published February 12, 2025 and initiated December 26, 2024. Recalling firm: ROi CPS LLC; Status: Ongoing; Reported quantity: 414 kits; Lot: Lot number 104006, Exp. 9/30/2026, UDI (01)10194717119074(17)260930(10)104006; and Lot number 103486, Exp. 8/31/2026, UDI (01)10194717119074(17)260831(10)103486. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is a potential for the Inline Controller to be using the spring out of specification causing a loss of positive pressure resulting in the Neo-Tee Resuscitator and/or Circuit will not reach the high and low ends of the PIP and PEEP pressure ranges impacting effective ventilation of the patient..
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- ROi CPS LLC
- Distributor
- Distribution was made to MO. There was no government/military/foreign distribution.
- Lot
- Lot number 104006, Exp. 9/30/2026, UDI (01)10194717119074(17)260930(10)104006; and Lot number 103486, Exp. 8/31/2026, UDI (01)10194717119074(17)260831(10)103486.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADefinium 656HD & Discovery XR656HD (GTINs: 00840682138598, 00195278840226, 00840682136709). X-Ray imagingFebruary 12, 2025
- FDASenTiva, REF: 1000C, Generators Part of the VNS Therapy System with Microburst Stimulation. Used for VagusFebruary 12, 2025
- FDAregard Clinical Packaging Solutions, HT00376I - SPECIAL PROCEDURE, Item Number 800223009September 25, 2024
- FDARegard Kit containing the Microtek decanterAugust 28, 2024
- FDAregard Operative LAP, Item Number 800943001; surgical convenience kitJanuary 24, 2024
- FDARegard Dressing change kit, Item Number 800026006January 10, 2024
Questions about this recall
Was regard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator. Convenience kit to be used in Labor and Delivery. recalled?
Yes. FDA published recall Z-1047-2025 on February 12, 2025. Reported hazard: There is a potential for the Inline Controller to be using the spring out of specification causing a loss of positive pressure resulting in the Neo-Tee Resuscitator and/or Circuit will not reach the high and low ends of the PIP and PEEP pressure ranges impacting effective ventilation of the patient..
What should I do if I have this product?
Voluntary: Firm initiated · Letter