AirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, Expiratory Filter, REF 2K8004F
Published March 6, 2024 · Recall initiated: January 10, 2024
Summary
FDA records medical device Class I recall Z-1059-2024, published March 6, 2024 and initiated January 10, 2024. Recalling firm: Vyaire Medical; Status: Ongoing; Reported quantity: 8,220 units; Lot: UDI/DI Case: 50190752114133, Each: 10190752114135; All manufacturing dates of 2017 and prior, and any products without a manufacturing date. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Vyaire Medical
- Distributor
- Worldwide distribution. US nationwide, Brazil, Canada, Switzerland, France, UK, Mexico, Netherlands, Saudi Arabia, Sweden, and South Africa
- Lot
- UDI/DI Case: 50190752114133, Each: 10190752114135; All manufacturing dates of 2017 and prior, and any products without a manufacturing date.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, Expiratory Filter, REFMarch 6, 2024
- FDAAirLife Adult Manual Resuscitator, Variable Volume Oxygen Reservoir Tubing, Adult Mask, REF 2K8017March 6, 2024
- FDAAirLife Adult Manual Resuscitator, 40" (1.0 m) Oxygen Reservoir Tubing, Adult Mask, CO2 Detector, REF 2K8005C2March 6, 2024
- FDAAirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, Adult Mask, CO2 Detector, REF 2K8004C2March 6, 2024
- FDAAirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, without Mask, REF 2K8001March 6, 2024
- FDAAirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, without Mask, REF 2K8000,March 6, 2024
Questions about this recall
Was AirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, Expiratory Filter, REF 2K8004F recalled?
Yes. FDA published recall Z-1059-2024 on March 6, 2024. Reported hazard: Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
What should I do if I have this product?
Voluntary: Firm initiated · Letter