Back to search
FDAFDA · Food and drugs·Recall number Z-1060-2024

AirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, Adult Mask, CO2 Detector, REF 2K8004C2

Published March 6, 2024 · Recall initiated: January 10, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Vyaire Medical
Distributor
Worldwide distribution. US nationwide, Brazil, Canada, Switzerland, France, UK, Mexico, Netherlands, Saudi Arabia, Sweden, and South Africa
Lot
UDI/DI Case: 50190752114119, Each: 10190752114111; All manufacturing dates of 2017 and prior, and any products without a manufacturing date.
Risk classification
Class I
Area (FDA)
device
Country
United States