AirLife Adult Manual Resuscitator, Variable Volume Oxygen Reservoir Tubing, Adult Mask, REF 2K8017
Published March 6, 2024 · Recall initiated: January 10, 2024
Summary
FDA records medical device Class I recall Z-1064-2024, published March 6, 2024 and initiated January 10, 2024. Recalling firm: Vyaire Medical; Status: Ongoing; Reported quantity: 1,020,330 units; Lot: UDI/DI Case: 50190752114294, Each: 10190752114296; All manufacturing dates of 2017 and prior, and any products without a manufacturing date. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Vyaire Medical
- Distributor
- Worldwide distribution. US nationwide, Brazil, Canada, Switzerland, France, UK, Mexico, Netherlands, Saudi Arabia, Sweden, and South Africa
- Lot
- UDI/DI Case: 50190752114294, Each: 10190752114296; All manufacturing dates of 2017 and prior, and any products without a manufacturing date.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, Expiratory Filter, REFMarch 6, 2024
- FDAAirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, Expiratory Filter, REF 2K8004FMarch 6, 2024
- FDAAirLife Adult Manual Resuscitator, 40" (1.0 m) Oxygen Reservoir Tubing, Adult Mask, CO2 Detector, REF 2K8005C2March 6, 2024
- FDAAirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, Adult Mask, CO2 Detector, REF 2K8004C2March 6, 2024
- FDAAirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, REF 2K8004March 6, 2024
- FDAVOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REFMarch 6, 2024
Questions about this recall
Was AirLife Adult Manual Resuscitator, Variable Volume Oxygen Reservoir Tubing, Adult Mask, REF 2K8017 recalled?
Yes. FDA published recall Z-1064-2024 on March 6, 2024. Reported hazard: Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
What should I do if I have this product?
Voluntary: Firm initiated · Letter