Juno DRF; Model: 709020 Multi-function x-ray systems capable of routine radiograph and general fluoroscopy procedures.
Published February 1, 2017 · Recall initiated: February 19, 2016
Summary
FDA records medical device Class II recall Z-1063-2017, published February 1, 2017 and initiated February 19, 2016. Recalling firm: Villa Radiology Systems LLC; Status: Ongoing; Reported quantity: 66 distributed in US; Lot: 10110770 11030805 11030806 11070845 11070848 11070849 11090864 11110900 12020978 12020982 12030999 12041003 12061014 12061018 12061020 12081040 12091074 12101104 12121186 12121187 13031215 13101263 14021301 14031313 14041321 14051332 14071353 14081359 14091368 14091371 14121400 14121403 15021426 15031440 15041444 15091473 15091478 15101484 15101487. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The analysis showed that the incident was caused by a chain of four faults: 1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position. 2) The directions for use (device labeling and instructions for use) did not contain an explicit exclusion of the prohibited area. 3.) The movement was activated inadvertently and continuously. and it was not released (the movement requires continuous activation). 4.) Delayed activation of the emergency stop.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Villa Radiology Systems LLC
- Distributor
- AL AZ CA CO CT DC FL IA IL IN LA MA MD MI MN MO MS NJ NY OH PA Puerto Rico TN TX UT WI
- Lot
- 10110770 11030805 11030806 11070845 11070848 11070849 11090864 11110900 12020978 12020982 12030999 12041003 12061014 12061018 12061020 12081040 12091074 12101104 12121186 12121187 13031215 13101263 14021301 14031313 14041321 14051332 14071353 14081359 14091368 14091371 14121400 14121403 15021426 15031440 15041444 15091473 15091478 15101484 15101487
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAApollo: Models: 9784000131, 9784000231, 9784020131, 9784020231, 9784100231, 9784120131, 9784120231, 9784130131February 1, 2017
- FDAApollo DRF: Model: 9784220831 Multi-function x-ray systems capable of routine radiograph and generalFebruary 1, 2017
- FDAApollo EZ: Model: 9784152036 Multi-function x-ray systems capable of routine radiograph and generalFebruary 1, 2017
- FDAVelara Generator with an old Firmware version used with the Philips Integris BH5000February 1, 2017
- FDAOEC 9900 Elite. MDL Numbers: D148942, D155043February 1, 2017
- FDAOwandy Radiology I -MAX Touch 3D X-rayNovember 18, 2015
Questions about this recall
Was Juno DRF; Model: 709020 Multi-function x-ray systems capable of routine radiograph and general fluoroscopy procedures. recalled?
Yes. FDA published recall Z-1063-2017 on February 1, 2017. Reported hazard: The analysis showed that the incident was caused by a chain of four faults: 1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position. 2) The directions for use (device labeling and instructions for use) did not contain an explicit exclusion of the prohibited area. 3.) The movement was activated inadvertently and continuously. and it was not released (the movement requires continuous activation). 4.) Delayed activation of the emergency stop.
What should I do if I have this product?
Voluntary: Firm initiated · Letter