Back to search
FDAFDA · Food and drugs·Recall number Z-1108-2026

Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 15.2.0, 15.1.3 Software Version: RayStation 2024A SP1, RayStation 2024A SP2 Product Description: Radiation Therapy Treatment Planning System Component: No

Published January 28, 2026 · Recall initiated: November 28, 2025

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
RAYSEARCH LABORATORIES AB
Distributor
Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS
Lot
UDI: 0735000201076120240508, 0735000201080820241217 GTIN: 07350002010761, 07350002010808 Serial Numbers: 15.1.0.852, 15.2.0.107 Software Revision: RayStation 2024A SP1, RayStation 2024A SP2 Expiration Date: 2030-04-15
Risk classification
Class II
Area (FDA)
device
Country
Sweden