FDAFDA · Food and drugs·Recall number Z-1115-2013

Product is distributed by Zimmer (Warsaw, Indiana)/Synvasive Technology, Inc. (El Dorado Hills, CA) & manufactured under Synvasive Technology, Inc. El Dorado Hills, CA Reciprocating saw blade: A sterile surgical saw blade used to resect damaged cartilage and/or bond in an orthopaedic reconstructive procedure such as a total knee replacement.

Published April 24, 2013 · Recall initiated: April 2, 2013

Summary

FDA records medical device Class II recall Z-1115-2013, published April 24, 2013 and initiated April 2, 2013. Recalling firm: Synvasive Technology Inc; Status: Terminated; Reported quantity: 3442 units per the part numbers listed; Lot: Lots Numbers included in Recall: 1X311,01251,16325,15023,15024,1Y224,23326,0X401,26385,1Y174, 17151,0Y280,17150,22028,28114,19071,23372,23346,23504,23505, 25134,25341,2W223,2X105,26001,2W573,2W574,24348,25304, 25333,26068,26165,27015,27016,29392,PD4647. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Various reciprocating and oscillating saw blades were incorrectly labeled. The labeling errors include: (1) The pouch says oscillating saw blade when it actually contains a reciprocating saw blade; (2) The box bar code date does not match the human readable expiration date; and (3) The label incorrectly states use with Zimmer/Synthes handset and not the Zimmer Universal Power System.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Synvasive Technology Inc
Distributor
Worldwide Distribution - USA (nationwide) and countries of: Austria, Australia, Belgium, Czech Republic, Denmark, Finland, France, Germany, Italy, Japan, Netherlands, Norway, Panama, Spain, Sweden, Switzerland, Tunisia, United Kingdom and Yemen.
Lot
Lots Numbers included in Recall: 1X311,01251,16325,15023,15024,1Y224,23326,0X401,26385,1Y174, 17151,0Y280,17150,22028,28114,19071,23372,23346,23504,23505, 25134,25341,2W223,2X105,26001,2W573,2W574,24348,25304, 25333,26068,26165,27015,27016,29392,PD4647
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Product is distributed by Zimmer (Warsaw, Indiana)/Synvasive Technology, Inc. (El Dorado Hills, CA) & manufactured under Synvasive Technology, Inc. El Dorado Hills, CA Reciprocating saw blade: A sterile surgical saw blade used to resect damaged cartilage and/or bond in an orthopaedic reconstructive procedure such as a total knee replacement. recalled?

Yes. FDA published recall Z-1115-2013 on April 24, 2013. Reported hazard: Various reciprocating and oscillating saw blades were incorrectly labeled. The labeling errors include: (1) The pouch says oscillating saw blade when it actually contains a reciprocating saw blade; (2) The box bar code date does not match the human readable expiration date; and (3) The label incorrectly states use with Zimmer/Synthes handset and not the Zimmer Universal Power System.

What should I do if I have this product?

Voluntary: Firm initiated · Letter