SR Instruments FG3105 Patient Scales Used on various Joerns Healthcare Patient Lifters. Model# FG3105 Patient Scales. Product Usage: The FG3105 Scale is designed for Joerns Healthcare for use in portable patient weighing systems. The scale is designed to be connected between the patient hoist mechanism and patient sling apparatus. The scale is not intended to be used for patient transport.
Published April 24, 2013 · Recall initiated: December 13, 2012
Summary
FDA records medical device Class II recall Z-1128-2013, published April 24, 2013 and initiated December 13, 2012. Recalling firm: Sr Instruments Inc; Status: Terminated; Reported quantity: 806 patient scales; Lot: Serial numbers: 1005 thru 1780, 1785 thru 1809, 1866 thru 1868 and 1870 thru 1875. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
SR Instruments Inc. is implementing a field correction for the Joerns Healthcare Patient Lifts after they received two incident reports regarding two field failures of a tensile type load cell involving the FG3105 Patient Scale. The failures resulted in patient falls.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Sr Instruments Inc
- Distributor
- USA Nationwide Distribution in the state of Texas.
- Lot
- Serial numbers: 1005 thru 1780, 1785 thru 1809, 1866 thru 1868 and 1870 thru 1875.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAThe Oxford Partial Knee; Oxford Knee Resection Procedure 3 Pack CEMENTED For use with Stryker System 5, SystemApril 24, 2013
- FDALifeScan brand OneTouch® Ultra Control Solution, Part Numbers/Model Numbers: 01045808, 02141603. Testing ofApril 24, 2013
- FDAAccuray CyberKnife System Medical Physicist, Robotic Radiosurgery System, Medical charged-particle radiationApril 24, 2013
- FDALifeScan brand OneTouch® Select Control Solution, Part Number/Model #: 02168902. Testing of Blood GlucoseApril 24, 2013
- FDAInstrumentation Industries BE183-SU and BE 183 SU-12 Manual Jet Ventilators; BE 183-SUR and BE 183 SUR-12April 24, 2013
- FDAHeritage Labs Path Study Extra Supplies Kit, Product # A 2376, kit contains 3 zippered plastic bags which holdApril 24, 2013
Questions about this recall
Was SR Instruments FG3105 Patient Scales Used on various Joerns Healthcare Patient Lifters. Model# FG3105 Patient Scales. Product Usage: The FG3105 Scale is designed for Joerns Healthcare for use in portable patient weighing systems. The scale is designed to be connected between the patient hoist mechanism and patient sling apparatus. The scale is not intended to be used for patient transport. recalled?
Yes. FDA published recall Z-1128-2013 on April 24, 2013. Reported hazard: SR Instruments Inc. is implementing a field correction for the Joerns Healthcare Patient Lifts after they received two incident reports regarding two field failures of a tensile type load cell involving the FG3105 Patient Scale. The failures resulted in patient falls.
What should I do if I have this product?
Voluntary: Firm initiated · Letter