LifeScan brand OneTouch® Select Control Solution, Part Number/Model #: 02168902. Testing of Blood Glucose Levels.

Published April 24, 2013 · Recall initiated: February 25, 2013

Summary

FDA records medical device Class II recall Z-1130-2013, published April 24, 2013 and initiated February 25, 2013. Recalling firm: Lifescan Inc; Status: Terminated; Reported quantity: 132,301 units total; Lot: 2AA2101, Exp. Date: 9-30-2014; 2AA2102, Exp. Date: 10-31-2-14. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

When OneTouch® Ultra and OneTouch® Select test strips are tested with Ultra and Select control solution the meter may display a result outside the control solution test range printed on the strip vial label even though the system is working properly.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Lifescan Inc
Distributor
Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, 10, IL, IN, KS, KY, LA, MA, MD, ME, Ml, MN, MO, MS, MT, NC, NO, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SO, TN, TX, UT, VA, VI, VT, WA, WI,WV, and WY, and the countries of Russia and Ukraine.
Lot
2AA2101, Exp. Date: 9-30-2014; 2AA2102, Exp. Date: 10-31-2-14
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was LifeScan brand OneTouch® Select Control Solution, Part Number/Model #: 02168902. Testing of Blood Glucose Levels. recalled?

Yes. FDA published recall Z-1130-2013 on April 24, 2013. Reported hazard: When OneTouch® Ultra and OneTouch® Select test strips are tested with Ultra and Select control solution the meter may display a result outside the control solution test range printed on the strip vial label even though the system is working properly.

What should I do if I have this product?

Voluntary: Firm initiated · Letter