FDAFDA · Food and drugs·Recall number Z-1163-2014

GE Healthcare, Innova 2100IQ, Innova 3100, Innova 3100IQ, Innova 4100, Innova 4100IQ Innova 2121IQ, Innova 3131IQ; Cardiovascular X-ray imaging systems (X-ray generator installed with Rotor V3 version and Heater V4 version boards). The Digital Fluoroscopic Imaging System is indicated for use in generating fluoroscopic images of human anatomy for diagnostic and intervention angiography procedures. It is intended to replace fluoroscopic images obtained through the image intensifier technology. This device is not intended for mammography applications. The Innova 3100 system is indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, and optionally, rotational and angiography procedures. It is intended to replace fluoroscopic images obtained through image intensifier technology. The Digital Fluoroscopic Imaging System is indicated for use in diagnostic and interventional angiographic procedures of human anatomy. It is intended to replace image intensifier fluoroscopic systems in all diagnostic or interventional procedures. This device is not intended for mammography applications. A new tilt table will support performing procedures like CO2 studies, Venography. The Digital Fluoroscopic Imaging System with Bolus Chasing option is indicated for use in generating fluoroscopic images of human anatomy for diagnostic and intervention; the option Bolus enables to perform subtraction angiography of lower limbs. This device is not intended for mammography applications. The Innova 2100-IQ system is indicated for use in generating fluoroscopic images of human anatomy for vascular angiography diagnostic and interventional procedures, and optionally, rotational angiography procedures. It is also indicated for generating fluoroscopic images of human anatomy for cardiology diagnostic and interventional procedures. It is intended to replace fluoroscopic images obtained through image intensifier technology. This device is not intended for mammography applications. For Innova 4100, Innova 4100IQ, Innova 3100, Innova 3100 IQ, Innova 2100IQ with Innova 3T Option: The Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures. They are also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures. They are intended to replace fluoroscopic images obtained through image intensifier technology. Those devices are not intended for mammography applications. Innova CT is a software option, which reconstructs 3D volumes from Rotational Fluoroscopy acquisition to provide images that assist the physician in diagnosis, surgical planning, interventional procedures and treatment follow-up. Innova CT is intended for imaging bone and soft tissues as well as other internal body structures. Innova CT is not intended for mammography applications. For Innova 4100, Innova 4100IQ, Innova 3100, Innova 3100 IQ, Innova 2100IQ with Innova 3D or InnovaSpin option: The Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures. They are also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures. They are intended to replace fluoroscopic images obtained through image intensifier technology. Those devices are not intended for mammography applications. Innova 3D is a software option which reconstructs 3D volumes from Rotational Fluoroscopy acquisition to assist the physician in diagnosis, surgical planning, interventional procedures and treatment follow-up. It is not intended for mammography applications. InnovaSpin is a software option that permits fast spin rotational angiography. It is not intended for mammography applications. :

Published March 19, 2014 · Recall initiated: December 11, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

GE Healthcare has become aware of a potential safety issue involving the innova systems. The Innova system can unexpectedly stop delivering X-rays after a power-on or a reset cycle which may result in the loss of real-time interventional imaging. No injuries have been reported due to this issue.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
GE Healthcare, LLC
Distributor
Worldwide Distribution - USA (nationwide) including the states of AL, AK, AZ, AR, CA, CO, CT, FL, GA, IL, IN, KS, KY, LA, ME, MD, MA, MI, MS, MO, NV, NJ, NY, NC, ND, OH, OK OR, PA, PR, RI, SC, TN, TX, UT, VA,WA,WY and WI., and the countries of ARGENTINA, AUSTRALIA, CANADA, CHINA, CZECH REPUBLIC, DENMARK, EGYPT, FINLAND, FRANCE, GERMANY, HUNGARY, INDIA, IRELAND, ITALY, JAPAN, JORDAN, REPUBLIC OF KOREA, MALAYSIA, MEXICO, NEW ZEALAND, NORWAY, PHILIPPINES, POLAND, RUSSIA, SAUDI ARABIA, SPAIN, SRI LANKA, TURKEY and UNITED KINGDOM.
Lot
Mfg Lot or Serial # System ID 00000550134BU1 2035763100A 00000581573BU3 203932VA4100 00000523132BU9 205783BCL5 00000560743BU7 206598UWBP 00000446090BU3 207351YCL1 00000504213BU0 212305CATH1 00000437215BU7 214324INNOVA2 00000562417BU6 2198864100 00000093723VE5 229228INNOV21C 00000583484BU1 240566CATH2 00000478109BU2 251435ICV3 00000477896BU5 3016094100 00000441137BU7 3018774100 00000466659BU0 303788CL1 00000466984BU2 304424INNOVA 00000445301BU5 309691P3100 00000424103BU0 310423INNOVA1 00000579612BU3 313745IN1 00451726BU4BU8 316268INNOVAA 00000466810BU9 318449INNOVA2 00000482326BU6 3184734100 00000499386BU1 330375LAB2 00000445906BU1 334286BSP 00000551081BU3 336832MCCL6 00000464558BU6 352333CATH2 00000463833BU4 361980INNOVA2 00000427056BU7 401274IN41 00000585912BU9 404712SP11 00000585915BU2 405271P4100 00000593051BU6 405610MWC3100EP 00000449428BU2 405840OIC3100 00000541272BU1 407518EP 00000431502BU4 407518SP3 00000452126BU6 407SPECIALS 00000490443BU9 408559INV1 00000421698BU2 4103374100 00000461599BU3 410337INNOVA6 00000571721BU0 410368INNOVA1 00000442436BU2 4105434100 00000533867BU8 410554INNOVA1 00000548841BU6 410554INNOVA4 00000560325BU3 410614CV1 00000443912BU1 412623LAB3 00000454264BU3 419251LAB1 00000576989BU8 423493CL1 00000450075BU7 480821IN2 00000468193BU8 501257VALAB1 00000434782BU9 502587LAB4 FMI12204DUMM03 508383IN4100 00000581740BU8 509473RM1 00000459476BU8 5135844100 00000415698BU0 5136864100 00000479625BU6 516562INNOVA5A 00000460641BU4 516562INNOVA6 00000460969BU9 516562INNOVA7 00000441966BU9 516663INNOVA2 00000467462BU8 518525LAB1 00000467645BU8 540953INNOVA 00000513959BU7 570476INNOVA21 00000466830BU7 573256MHC 00000479758BU5 573815BCR2 00000433021BU3 573882UMC1 00000446439BU2 6012884100SP1 00000558894BU2 601376SP4100 00000437985BU5 6082634100 00000410733BU0 608324DAL 00000543076BU4 608775NOVA1 00000561529BU9 609267IN2100 00000546435BU9 609757IN3100 00000491706BU8 610250INNOVA3 00000442603BU7 614293LIBERTY2 00000563635BU2 614722INNOVA1 00000458814BU1 615338CV2 00000493727BU2 615446CV1 00000458501BU4 619229INNOVA1 00000438341BU0 626359INNOVA 00000463834BU2 65049331 00000546155BU3 7018573100 00000469683BU7 7028533100 00000433013BU0 706278INNOVA1 00000421893BU9 706802FINN 00000454140BU5 7083333100 00000427535BU0 713776M3100 00000555540BU4 717782INNOVA31B 00000459460BU2 718283INNOVA3 00000490237BU5 718470INNOVA2 00000436326BU3 718518CVL1 00000398531BU4 727725SPEC1 00000571258BU3 727942CATH 00000559817BU2 732776CATH2 00000533869BU4 732776CATH6 00000460652BU1 765453INNOVA1 00000467397BU6 770535N4100 00000478491BU4 770991CL3 00000449237BU7 801268INNOVA 00000432806BU8 80140841 00000479709BU8 804594CATH 00000438367BU5 804764CATH3 00000442737BU3 812283IN31 00000592458BU4 812450CL2 00000479578BU7 812858GWLAB2 00000492209BU2 813972INTV 00000439429BU2 815226INN2 00000434732BU4 815599FHN3100 00000538053BU0 8157412100 00000467323BU2 816478MC1 00000477463BU4 817329BG2100 00000557168BU2 8189813100 00000447651BU1 828298VASC 00000460469BU0 845365V19 00000523086BU7 845368LAB1 00000544815BU4 850469BCV1 00000544202BU5 850747IN4100 00000485433BU7 856596IN3100 00000491032BU9 858554INNOVA2 00000422738BU5 8595724100 00000467881BU9 864255SF4100 00000548598BU2 864255SFCL2 00000537821BU1 865693CCL1 00000596870BU6 903CSM2121 00000458184BU9 903GS4100 00000454594BU3 904202CATH3 00000496915BU0 907562CATH1 00000436762BU9 918744JR4100 00000549261BU6 9284532100 00000442333BU1 9408984100 00000439384BU9 951788INNOVA4 00000545588BU6 954987CL4 00000482458BU7 954MW4100 00000430321BU0 956632INN4100 00000469847BU8 956698INN31 00000430322BU8 985867INNOVA2 00000514612BU1 COR359867 00000403215BU7 FWANG3217 00000580430BU7 2036884100B 00000590065BU9 303399EP 00000571259BU1 318449INNOVA3 00000057472BU0 352333CV3 00000590868BU6 361788CATH1 00000543267BU9 410337INNOVA2 00000590608BU6 412623H 00000480416BU7 423493I4100 0000
Risk classification
Class II
Area (FDA)
device
Country
United States