10 cc Luer-Slip Loss of Resistance Syringe, Product Code LR-05501

Published February 19, 2020 · Recall initiated: January 15, 2020

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Arrow International Inc
Distributor
The products were distributed to the following US states: AK, AR, AZ, CA, CO, CT, FL, GA, ID, IL, IN, LA, MA, MD, MI, MN, MO, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WY.
Lot
Lots: 13F18J0916 13F19C0321 13F19G0070 13F19H0274 13F19J0039 13F19L0252 13F18K0675 13F19D0335 13F19G0071 13F19H0275 13F19J0040 13F18L0737 13F19E0141 13F19G0458 13F19H0634 13F19J0596 13F18M0178 13F19E0212 13F19H0085 13F19J0038 13F19K0097
Risk classification
Class II
Area (FDA)
device
Country
United States