FDAFDA · Food and drugs·Recall number Z-1226-2014
VITROS Chemistry Products White Reference Slides, White Correction Factor Slide (WCF), DT Slides, 25 Slides, Part Number J02315; Made in USA by Ortho- Clinical Diagnostics, Inc. 100 Indigo Creek Drive, Rochester, NY 14626 --- This product is a part/ tool used when performing the correction factors adjustment on the VITROS Chemistry Systems. It does not have a 510(k) number. Because the WCF slide is a part/ tool, and is not a device used for diagnostic testing, there is no Intended Use statement. The following summary is provided as a description/ explanation of the WCF slide. The WCF slide is a part/ tool used in the reflectometer correction factor adjustment procedure in order to optimize the optical calibration of the VITROS 250/350/5,1 FS/4600 Chemistry Systems and the VITROS 5600 Integrated
Published March 26, 2014 · Recall initiated: January 9, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
An incorrect assay value (0.3785) was printed on the product carton for the 460 nm wavelength (filter) that may cause a delay in the ability to obtain and report TBIL, Bu, Bc, and derived tests results. The correct assay value is 0.8658.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Ortho-Clinical Diagnostics
- Distributor
- Worldwide Distribution - USA (nationwide) Puerto Rico, and Internationally to Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Singapore, Spain, and Venezuela.
- Lot
- Lot 9052-0045-6358, expiry date 01 Aug 2015
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.