Back to search
FDAFDA · Food and drugs·Recall number Z-1227-2024

VITROS XT 3400 Chemistry System, Catalog No. 6844458

Published March 13, 2024 · Recall initiated: January 26, 2024

Summary

FDA records medical device Class II recall Z-1227-2024, published March 13, 2024 and initiated January 26, 2024. Recalling firm: Ortho-Clinical Diagnostics, Inc.; Status: Ongoing; Reported quantity: 7 US; 4 OUS; Lot: VITROS System Software Version 3.8.1 UDI-DI 10758750031986 Serial Numbers US: J34500180 J34500289 J34500422 J34500473 J34500622 J34500861 J34500887 OUS: J34500515 J34500609 J34500802 J34500940. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Ortho-Clinical Diagnostics, Inc.
Distributor
Worldwide - US Nationwide distribution in the states of AZ, CA, CT, FL, IA, ID, IL, IN, KY, LA, MO, NE, NY, OH, OR, SC, SD, TX, WA, WI, WV and the countries of Australia, Brazil, Canada, China, Colombia, India, Mexico, Singapore, United Kingdom, Italy, Spain, Russia, France, and Belgium.
Lot
VITROS System Software Version 3.8.1 UDI-DI 10758750031986 Serial Numbers US: J34500180 J34500289 J34500422 J34500473 J34500622 J34500861 J34500887 OUS: J34500515 J34500609 J34500802 J34500940
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was VITROS XT 3400 Chemistry System, Catalog No. 6844458 recalled?

Yes. FDA published recall Z-1227-2024 on March 13, 2024. Reported hazard: A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.

What should I do if I have this product?

Voluntary: Firm initiated · Letter