VITROS 3600 Immunodiagnostic System Catalog Nos. 6802783 (regular) and 6802914 (refurbished)
Published March 13, 2024 · Recall initiated: January 26, 2024
Summary
FDA records medical device Class II recall Z-1229-2024, published March 13, 2024 and initiated January 26, 2024. Recalling firm: Ortho-Clinical Diagnostics, Inc.; Status: Ongoing; Reported quantity: 9 US; 66 OUS; Lot: VITROS System Software Versions 3.8.0 or 3.8.1 Regular UDI-DI 10758750002979 Refurbished UD-DI 10758750007103 Serial Numbers: US: J36000109 J36000292 J36000489 J36000955 J36001167 J36001222 J36001283 J36001330 J36001348 OUS: J36000148 J36001000 J36001495 J36000340 J36001012 J36001513 J36000348 J36001024 J36001515 J36000406 J36001026 J36001519 J36000407 J36001027 J36001528 J36000427 J36001051 J36001530 J36000446 J36001088 J36001535 J36000511 J36001108 J36001558 J36000517 J36001111 J36001570 J36000580 J36001112 J36001573 J36000612 J36001117 J36001575 J36000626 J36001154 J36001579 J36000688 J36001229 J36001581 J36000753 J36001237 J36001582 J36000861 J36001245 J36001583 J36000877 J36001246 J36001595 J36000920 J36001401 J36001600 J36000923 J36001403 J36001602 J36000929 J36001427 J36001608 J36000933 J36001451 J36001618 J36000941 J36001488 J36001632 J36000942 J36001489 J36001640. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not flagged or reported with the applicable QC results if the statistical criteria is met. This may result in acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ortho-Clinical Diagnostics, Inc.
- Distributor
- Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
- Lot
- VITROS System Software Versions 3.8.0 or 3.8.1 Regular UDI-DI 10758750002979 Refurbished UD-DI 10758750007103 Serial Numbers: US: J36000109 J36000292 J36000489 J36000955 J36001167 J36001222 J36001283 J36001330 J36001348 OUS: J36000148 J36001000 J36001495 J36000340 J36001012 J36001513 J36000348 J36001024 J36001515 J36000406 J36001026 J36001519 J36000407 J36001027 J36001528 J36000427 J36001051 J36001530 J36000446 J36001088 J36001535 J36000511 J36001108 J36001558 J36000517 J36001111 J36001570 J36000580 J36001112 J36001573 J36000612 J36001117 J36001575 J36000626 J36001154 J36001579 J36000688 J36001229 J36001581 J36000753 J36001237 J36001582 J36000861 J36001245 J36001583 J36000877 J36001246 J36001595 J36000920 J36001401 J36001600 J36000923 J36001403 J36001602 J36000929 J36001427 J36001608 J36000933 J36001451 J36001618 J36000941 J36001488 J36001632 J36000942 J36001489 J36001640
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively measures C-reactive protein (CRP)December 4, 2024
- FDAThe VITROS XT Chemistry Products ALB-TP Slides is a single device that contains both an albumin test and aAugust 28, 2024
- FDAVITROS Immunodiagnostic VITROS Folate Reagent Pack 1/2August 21, 2024
- FDAVITROS Chemistry Products Calibrator Kit 29- In-Vitor used to calibrate the VITROS 5,1 FS/4600 ChemistryJuly 10, 2024
- FDAVITROS XT 7600 Integrated System, Catalog No. 6844461March 13, 2024
- FDAVITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)March 13, 2024
Questions about this recall
Was VITROS 3600 Immunodiagnostic System Catalog Nos. 6802783 (regular) and 6802914 (refurbished) recalled?
Yes. FDA published recall Z-1229-2024 on March 13, 2024. Reported hazard: A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not flagged or reported with the applicable QC results if the statistical criteria is met. This may result in acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported.
What should I do if I have this product?
Voluntary: Firm initiated · Letter