Guardian II Hemostasis Valve, models 8210 and 8211. The following models are only available Outside the US: FH101, FH101-T, FH101-25, and FH101-50. Product Usage: The Guardian hemostasis valves are designed to be used as conduits when interventional devices with diameters up to 8 F are inserted into the human vascular system. The devices have two seals: the low-pressure seal, and the high-pressure seal. Closure of the high-pressure seal, which is achieved when the nut can no longer rotate, secures the diagnostic/interventional device in position within the vasculature and allows for pressure injections up to 600 psi.
Published April 13, 2016 · Recall initiated: March 4, 2016
Summary
FDA records medical device Class I recall Z-1275-2016, published April 13, 2016 and initiated March 4, 2016. Recalling firm: Vascular Solutions, Inc.; Status: Terminated; Reported quantity: 5283; Lot: Lot numbers: 41776, 41777, 41874, 42108, 42525, 581766, 581767, 583285, and 583286. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Vascular Solutions became aware of a potential problem with the click version of the Guardian II hemostasis valves. The low pressure seal may not close properly, which may allow air to be introduced into the device and may lead to risk of an air embolism.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Vascular Solutions, Inc.
- Distributor
- Nationwide Distribution in the states of AL, AR, CA, CO, DE, FL, GA, ID, IL IN, IA,KY, LA, MD, MA, MI, MO, NE,NV, NH, NY, NC, OH, PA, SD, TN, TX, UT, VA, WA, and WV. OUS: Ireland
- Lot
- Lot numbers: 41776, 41777, 41874, 42108, 42525, 581766, 581767, 583285, and 583286.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADrainer(R) Centesis Catheters: (a) REF 8810, 5F x 9cm (b) REF 8811, 6F x 9cm (c) REF 8812, 5F x 15cm ProductJune 27, 2018
- FDADolphin Inflation Device, REF 0185NA. Rx only, Sterilized using Ethylene Oxide. DOLPHIN is a single-use,June 7, 2017
- FDAGrebSet Micro-Introducer Kit, model 7941, Sterilized with Ethylene Oxide. Product Usage: The GrebSetMarch 1, 2017
- FDAVascular Solutions Twin-Pass .023" Dual Access Catheter, model 5230. Sterilized with Ethylene Oxide. TheOctober 26, 2016
- FDAGE Inspection Technologies, LP, Cabinet X-ray system used for industrial quality control.April 13, 2016
- FDAAll instrumentation associated with the Prelude PF Resurfacing Knee System.April 13, 2016
Questions about this recall
Was Guardian II Hemostasis Valve, models 8210 and 8211. The following models are only available Outside the US: FH101, FH101-T, FH101-25, and FH101-50. Product Usage: The Guardian hemostasis valves are designed to be used as conduits when interventional devices with diameters up to 8 F are inserted into the human vascular system. The devices have two seals: the low-pressure seal, and the high-pressure seal. Closure of the high-pressure seal, which is achieved when the nut can no longer rotate, secures the diagnostic/interventional device in position within the vasculature and allows for pressure injections up to 600 psi. recalled?
Yes. FDA published recall Z-1275-2016 on April 13, 2016. Reported hazard: Vascular Solutions became aware of a potential problem with the click version of the Guardian II hemostasis valves. The low pressure seal may not close properly, which may allow air to be introduced into the device and may lead to risk of an air embolism.
What should I do if I have this product?
Voluntary: Firm initiated · Letter